NABL calibration scope CC-4643

Search the 228 electrical, mechanical and thermal calibration entries in Technex / RTS scope CC-4643 by instrument, range and laboratory or site facility.

Find a measurement in scope CC-4643

This directory follows the 228 entries in the published scope for Technex Calibration (OPC) Private Limited: 133 at the permanent facility and 95 at customer sites. Search for an instrument or quantity, then compare its range, method and facility with the work you need. Entry numbers restart in the site-facility section.

CMC means calibration and measurement capability. The annexure expresses these values as expanded uncertainties at approximately 95% confidence, with coverage factor k = 2. The result reported for a particular instrument also depends on the conditions of its calibration. Each entry links to its page in the original scope.

First page of the actual scope of accreditation CC-4643Download the complete scope45-page PDF · original document

Certificate and catalogue

228 scope entries

CC-4643 · ISO/IEC 17025:2017 · validity 9 January 2026 to 8 January 2030
Instrument / quantityRange and conditionsMethodCMC (±)Source
AC Current @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Measure)Permanent facility · entry 11 A to 10 AUsing 5½ Digital Multimeter by Direct Method1.28 %Page 1
AC Current @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Measure)Permanent facility · entry 210 mA to 100 mAUsing 5½ Digital Multimeter by Direct Method0.41 % to 0.43 %Page 1
AC Current @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Measure)Permanent facility · entry 3100 mA to 200 mAUsing 5½ Digital Multimeter by Direct Method0.43 % to 0.38 %Page 1
AC Current @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Measure)Permanent facility · entry 42 mA to 10 mAUsing 5½ Digital Multimeter by Direct Method1.31 % to 0.41 %Page 1
AC Voltage @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Measure)Permanent facility · entry 51 V to 10 VUsing 5½ Digital Multimeter by Direct Method0.31 %Page 2
AC Voltage @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Measure)Permanent facility · entry 610 V to 100 VUsing 5½ Digital Multimeter by Direct Method0.31 % to 0.30 %Page 2
AC Voltage @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Measure)Permanent facility · entry 7100 mV to 200 mVUsing 5½ Digital Multimeter by Direct Method0.30 % to 0.24 %Page 2
AC Voltage @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Measure)Permanent facility · entry 8100 V to 500 VUsing 5½ Digital Multimeter by Direct Method0.30 % to 0.28 %Page 2
AC Voltage @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Measure)Permanent facility · entry 920 mV to 100 mVUsing 5½ Digital Multimeter by Direct Method0.79 % to 0.30 %Page 2
AC Voltage @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Measure)Permanent facility · entry 10500 V to 750 VUsing 5½ Digital Multimeter by Direct Method0.28 % to 0.31 %Page 3
AC Current @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Source)Permanent facility · entry 110.2 mA to 1 mAUsing 5½ Multifunction Calibrator by Direct Method0.64 % to 0.26 %Page 3
AC Current @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Source)Permanent facility · entry 121 A to 10 AUsing 5½ Multifunction Calibrator by Direct Method0.36 % to 0.22 %Page 3
AC Current @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Source)Permanent facility · entry 131 mA to 20 mAUsing 5½ Multifunction Calibrator by Direct Method0.26 % to 0.27 %Page 3
AC Current @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Source)Permanent facility · entry 14100 mA to 200 mAUsing 5½ Multifunction Calibrator by Direct Method0.20 % to 0.027 %Page 3
AC Current @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Source)Permanent facility · entry 1520 mA to 100 mAUsing 5½ Multifunction Calibrator by Direct Method0.27 % to 0.20 %Page 4
AC Current @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Source)Permanent facility · entry 16200 mA to 2000 mAUsing 5½ Multifunction Calibrator by Direct Method0.027 % to 0.42 %Page 4
AC Current @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Source)Permanent facility · entry 1710 A to 700 AUsing 5½ Multifunction Calibrator with Current Coil by Direct Method2.05 % to 0.97 %Page 4
AC Voltage @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Source)Permanent facility · entry 181 V to 10 VUsing 5½ Multifunction Calibrator by Direct Method0.97 % to 0.44 %Page 4
AC Voltage @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Source)Permanent facility · entry 1910 mV to 100 mVUsing 5½ Multifunction Calibrator by Direct Method0.24 % to 0.19 %Page 4
AC Voltage @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Source)Permanent facility · entry 2010 V to 20 VUsing 5½ Multifunction Calibrator by Direct Method0.44 % to 0.33 %Page 4
AC Voltage @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Source)Permanent facility · entry 21100 mV to 200 mVUsing 5½ Multifunction Calibrator by Direct Method0.19 % to 0.21 %Page 5
AC Voltage @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Source)Permanent facility · entry 22100 V to 200 VUsing 5½ Multifunction Calibrator by Direct Method0.20 % to 0.19 %Page 5
AC Voltage @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Source)Permanent facility · entry 2320 V to 100 VUsing 5½ Multifunction Calibrator by Direct Method0.33 % to 0.20 %Page 5
AC Voltage @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Source)Permanent facility · entry 24200 V to 500 VUsing 5½ Multifunction Calibrator by Direct Method0.19 % to 0.21 %Page 5
AC Voltage @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Source)Permanent facility · entry 25500 V to 900 VUsing 5½ Multifunction Calibrator by Direct Method0.21 % to 0.18 %Page 5
AC Voltage @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Source)Permanent facility · entry 26900 V to 1000 VUsing 5½ Multifunction Calibrator by Direct Method0.18 % to 0.20 %Page 5
DC CurrentELECTRO- TECHNICAL- DIRECT CURRENT (Measure)Permanent facility · entry 271 A to 10 AUsing 5½ Digital Multimeter by Direct Method0.13 % to 0.28 %Page 6
DC CurrentELECTRO- TECHNICAL- DIRECT CURRENT (Measure)Permanent facility · entry 281 mA to 10 mAUsing 5½ Digital Multimeter by Direct Method0.62 % to 0.084 %Page 6
DC CurrentELECTRO- TECHNICAL- DIRECT CURRENT (Measure)Permanent facility · entry 2910 mA to 100 mAUsing 5½ Digital Multimeter by Direct Method0.084 % to 0.078 %Page 6
DC CurrentELECTRO- TECHNICAL- DIRECT CURRENT (Measure)Permanent facility · entry 30100 mA to 200 mAUsing 5½ Digital Multimeter by Direct Method0.078 % to 0.076 %Page 6
DC CurrentELECTRO- TECHNICAL- DIRECT CURRENT (Measure)Permanent facility · entry 31200 mA to 1 AUsing 5½ Digital Multimeter by Direct Method0.076 % to 0.13 %Page 6
DC VoltageELECTRO- TECHNICAL- DIRECT CURRENT (Measure)Permanent facility · entry 321 mV to 200 mVUsing 5½ Digital Multimeter by Direct Method0.71 % to 0.017 %Page 6
DC VoltageELECTRO- TECHNICAL- DIRECT CURRENT (Measure)Permanent facility · entry 331 V to 1000 VUsing 5½ Digital Multimeter by Direct Method0.061 % to 0.02 %Page 7
DC VoltageELECTRO- TECHNICAL- DIRECT CURRENT (Measure)Permanent facility · entry 34200 mV to 1 VUsing 5½ Digital Multimeter by Direct Method0.017 % to 0.061 %Page 7
Resistance (2 Wire)ELECTRO- TECHNICAL- DIRECT CURRENT (Measure)Permanent facility · entry 351 kohm to 10 kohmUsing 5½ Digital Multimeter by Direct Method0.030 % to 0.026 %Page 7
Resistance (2 Wire)ELECTRO- TECHNICAL- DIRECT CURRENT (Measure)Permanent facility · entry 361 Mohm to 10 MohmUsing 5½ Digital Multimeter by Direct Method0.13 % to 0.27 %Page 7
Resistance (2 Wire)ELECTRO- TECHNICAL- DIRECT CURRENT (Measure)Permanent facility · entry 3710 kohm to 100 kohmUsing 5½ Digital Multimeter by Direct Method0.026 % to 0.027 %Page 7
Resistance (2 Wire)ELECTRO- TECHNICAL- DIRECT CURRENT (Measure)Permanent facility · entry 3810 Mohm to 90 MohmUsing 5½ Digital Multimeter by Direct Method0.27 % to 4.72 %Page 7
Resistance (2 Wire)ELECTRO- TECHNICAL- DIRECT CURRENT (Measure)Permanent facility · entry 39100 kohm to 1 MohmUsing 5½ Digital Multimeter by Direct Method0.027 % to 0.13 %Page 8
Resistance (4 Wire)ELECTRO- TECHNICAL- DIRECT CURRENT (Measure)Permanent facility · entry 4020 ohm to 180 ohmUsing 5½ Digital Multimeter by Direct Method0.700 % to 0.023 %Page 8
DC CurrentELECTRO- TECHNICAL- DIRECT CURRENT (Source)Permanent facility · entry 410.1 mA to 1 mAUsing 5½ Multifunction Calibrator by Direct Method0.7 % to 0.13 %Page 8
DC CurrentELECTRO- TECHNICAL- DIRECT CURRENT (Source)Permanent facility · entry 421 A to 10 AUsing 5½ Multifunction Calibrator by Direct Method0.24 % to 0.16 %Page 8
DC CurrentELECTRO- TECHNICAL- DIRECT CURRENT (Source)Permanent facility · entry 431 mA to 2 mAUsing 5½ Multifunction Calibrator by Direct Method0.13 %Page 8
DC CurrentELECTRO- TECHNICAL- DIRECT CURRENT (Source)Permanent facility · entry 4410 A to 900 AUsing 5½ Multifunction Calibrator by Direct Method1.91 % to 0.90 %Page 8
DC CurrentELECTRO- TECHNICAL- DIRECT CURRENT (Source)Permanent facility · entry 452 mA to 20 mAUsing 5½ Multifunction Calibrator by Direct Method0.16 % to 0.071 %Page 9
DC CurrentELECTRO- TECHNICAL- DIRECT CURRENT (Source)Permanent facility · entry 4620 mA to 200 mAUsing 5½ Multifunction Calibrator by Direct Method0.071 % to 0.14 %Page 9
DC CurrentELECTRO- TECHNICAL- DIRECT CURRENT (Source)Permanent facility · entry 47200 mA to 1 AUsing 5½ Multifunction Calibrator by Direct Method0.140 % to 0.24 %Page 9
DC VoltageELECTRO- TECHNICAL- DIRECT CURRENT (Source)Permanent facility · entry 480.2 V to 1000 VUsing 5½ Multifunction Calibrator by Direct Method0.18 % to 0.13 %Page 9
DC VoltageELECTRO- TECHNICAL- DIRECT CURRENT (Source)Permanent facility · entry 491 mV to 200 mVUsing 5½ Multifunction Calibrator By Direct Method1.40 % to 0.13 %Page 9
Resistance (2 Wire)ELECTRO- TECHNICAL- DIRECT CURRENT (Source)Permanent facility · entry 501 kohm to 10 kohmUsing Resistance Decade box by Direct Method0.14 % to 0.12 %Page 9
Resistance (2 Wire)ELECTRO- TECHNICAL- DIRECT CURRENT (Source)Permanent facility · entry 511 Mohm to 10 MohmUsing Resistance Decade box by Direct Method0.12 % to 3.52 %Page 10
Resistance (2 Wire)ELECTRO- TECHNICAL- DIRECT CURRENT (Source)Permanent facility · entry 5210 kohm to 100 kohmUsing Resistance Decade box by Direct Method0.12 %Page 10
Resistance (2 Wire)ELECTRO- TECHNICAL- DIRECT CURRENT (Source)Permanent facility · entry 5310 Mohm to 100 MohmUsing Decade Resistance Box by Direct Method3.52 % to 2.40 %Page 10
Resistance (2 Wire)ELECTRO- TECHNICAL- DIRECT CURRENT (Source)Permanent facility · entry 54100 kohm to 1 MohmUsing Resistance Decade box by Direct Method0.12 %Page 10
Resistance (2 Wire)ELECTRO- TECHNICAL- DIRECT CURRENT (Source)Permanent facility · entry 55100 ohm to 1 kohmUsing Resistance Decade box by Direct Method0.12 % to 0.14 %Page 10
Resistance (4 Wire)ELECTRO- TECHNICAL- DIRECT CURRENT (Source)Permanent facility · entry 5610 ohm to 100 ohmUsing Resistance Decade Box by Direct Method0.13 % to 0.12 %Page 10
B Type ThermocoupleELECTRO- TECHNICAL- TEMPERATURE SIMULATION (Source)Permanent facility · entry 57600 °C to 1800 °CUsing Universal calibrator by Direct Method3.45 °CPage 11
E Type ThermocoupleELECTRO- TECHNICAL- TEMPERATURE SIMULATION (Source)Permanent facility · entry 58(-) 200 °C to 1000 °CUsing Universal Calibrator by Direct Method2.90 °CPage 11
J Type ThermocoupleELECTRO- TECHNICAL- TEMPERATURE SIMULATION (Source)Permanent facility · entry 59(-) 200 °C to 1200 °CUsing Universal Calibrator by Direct Method3.08 °CPage 11
K Type ThermocoupleELECTRO- TECHNICAL- TEMPERATURE SIMULATION (Source)Permanent facility · entry 60(-) 200 °C to 1370 °CUsing Universal Calibrator by Direct Method3.19 °CPage 11
N Type ThermocoupleELECTRO- TECHNICAL- TEMPERATURE SIMULATION (Source)Permanent facility · entry 61(-) 200 °C to 1300 °CUsing Universal Calibrator by Direct Method2.24 °CPage 11
R Type ThermocoupleELECTRO- TECHNICAL- TEMPERATURE SIMULATION (Source)Permanent facility · entry 620 °C to 1750 °CUsing Universal Calibrator by Direct Method3.38 °CPage 11
RTD (Pt 100)ELECTRO- TECHNICAL- TEMPERATURE SIMULATION (Source)Permanent facility · entry 63(-) 200 °C to 800 °CUsing Universal Calibrator by Direct Method1.32 °CPage 12
S Type ThermocoupleELECTRO- TECHNICAL- TEMPERATURE SIMULATION (Source)Permanent facility · entry 640 °C to 1750 °CUsing Universal Calibrator by Direct Method3.54 °CPage 12
T Type ThermocoupleELECTRO- TECHNICAL- TEMPERATURE SIMULATION (Source)Permanent facility · entry 65(-) 200 °C to 400 °CUsing Universal Calibrator by Direct Method2.31 °CPage 12
FrequencyELECTRO- TECHNICAL- TIME & FREQUENCY (Measure)Permanent facility · entry 6645 Hz to 500 HzUsing 5½ Digital Multimeter by Direct Method0.48 % to 0.034 %Page 12
FrequencyELECTRO- TECHNICAL- TIME & FREQUENCY (Measure)Permanent facility · entry 67500 Hz to 1000 HzUsing 5½ Digital Multimeter by Direct Method0.034 % to 0.063 %Page 12
TimeELECTRO- TECHNICAL- TIME & FREQUENCY (Measure)Permanent facility · entry 680.1 s to 60 sUsing Time Calibration Standard by Direct Method0.01 s to 0.03 sPage 12
TimeELECTRO- TECHNICAL- TIME & FREQUENCY (Measure)Permanent facility · entry 693600 s to 86400 sUsing Time Calibration Standard by Direct Method0.79 s to 5.58 sPage 13
TimeELECTRO- TECHNICAL- TIME & FREQUENCY (Measure)Permanent facility · entry 7060 s to 3600 sUsing Time Calibration Standard by Direct Method0.03 s to 0.79 sPage 13
FrequencyELECTRO- TECHNICAL- TIME & FREQUENCY (Source)Permanent facility · entry 7145 Hz to 500 HzUsing 5½ Multifunction Calibrator by Direct Method0.30 % to 0.026 %Page 13
FrequencyELECTRO- TECHNICAL- TIME & FREQUENCY (Source)Permanent facility · entry 72500 Hz to 1000 HzUsing 5½ Multifunction Calibrator by Direct Method0.026 % to 0.02 %Page 13
Digital Tachometer (Non Contact-Type)MECHANICAL- ACCELERATION AND SPEEDPermanent facility · entry 73> 100 rpm to 1000 rpmUsing Digital Tachometer with RPM Source by Comparison Method4.01 % of rdgPage 13
Digital Tachometer (Non Contact-Type)MECHANICAL- ACCELERATION AND SPEEDPermanent facility · entry 74> 1000 rpm to 90000 rpmUsing Digital Tachometer with RPM Source by Comparison Method4.00 % of rdgPage 13
Digital Tachometer (Non Contact-Type)MECHANICAL- ACCELERATION AND SPEEDPermanent facility · entry 7510 rpm to 100 rpmUsing Digital Tachometer with RPM Source by Comparison Method5.6 % of rdgPage 14
Analog/Dial/Digital External Micrometer - L.C.: 0.001 mm & CoarserMECHANICAL- DIMENSION (BASIC MEASURING INSTRUMENT, GAUGE ETC.)Permanent facility · entry 76> 25 mm to 75 mmUsing Slip Gauge Set by Comparison Method1.52 µmPage 14
Analog/Dial/Digital External Micrometer - L.C.: 0.001 mm & CoarserMECHANICAL- DIMENSION (BASIC MEASURING INSTRUMENT, GAUGE ETC.)Permanent facility · entry 770 to 25 mmUsing Slip Gauge Set by Comparison Method1.49 µmPage 14
Analog/Dial/Digital Height Gauge - L.C.: 0.01 mm & CoarserMECHANICAL- DIMENSION (BASIC MEASURING INSTRUMENT, GAUGE ETC.)Permanent facility · entry 780 to 600 mmUsing Slip Gauge Set, Surface Plate & Caliper Checker by Comparison Method11.95 µmPage 14
Analog/Dial/Digital Vernier Caliper - L.C.: 0.01 mm & CoarserMECHANICAL- DIMENSION (BASIC MEASURING INSTRUMENT, GAUGE ETC.)Permanent facility · entry 790 to 150 mmUsing Slip Gauge Set, Caliper Checker by Comparison Method13.36 µmPage 14
Analog/Dial/Digital Vernier Caliper - L.C.: 0.01 mm & CoarserMECHANICAL- DIMENSION (BASIC MEASURING INSTRUMENT, GAUGE ETC.)Permanent facility · entry 800 to 300 mmUsing Slip Gauge Set, Caliper Checker by Comparison Method13.36 µmPage 15
Analog/Dial/Digital Vernier Caliper - L.C.: 0.01 mm & CoarserMECHANICAL- DIMENSION (BASIC MEASURING INSTRUMENT, GAUGE ETC.)Permanent facility · entry 810 to 600 mmUsing Slip Gauge Set, Caliper Checker by Comparison Method13.06 µmPage 15
Analog/Digital Coating Thickness Gauge - L.C.: 0.001 mm & CoarserMECHANICAL- DIMENSION (BASIC MEASURING INSTRUMENT, GAUGE ETC.)Permanent facility · entry 820.05 mm to 2 mmStandard Thickness Foils by Comparison Method3.06 µmPage 15
Angle Protractor - L.C.: 1°MECHANICAL- DIMENSION (BASIC MEASURING INSTRUMENT, GAUGE ETC.)Permanent facility · entry 830° to 90°Using Angle VMM (Vision Measuring Machine) by Comparison Method126.26 Seconds of ArcPage 15
Bevel Protector L.C.: 5'MECHANICAL- DIMENSION (BASIC MEASURING INSTRUMENT, GAUGE ETC.)Permanent facility · entry 840° to 180°Using Angle Gauge Block Set by Comparison Method1.19 Minutes of ArcPage 15
Bore Gauge - (Transmission) - L.C : 0.001 mmMECHANICAL- DIMENSION (BASIC MEASURING INSTRUMENT, GAUGE ETC.)Permanent facility · entry 85Up to 2 mmUsing Dial Calibration Tester by Comparison Method1.43 µmPage 16
Combination Set - L.C.: 1°MECHANICAL- DIMENSION (BASIC MEASURING INSTRUMENT, GAUGE ETC.)Permanent facility · entry 860° to 180°Using Angle Gauge Block Set / VMM (Vision Measuring Machine) by Comparison Method126.18 Seconds of ArcPage 16
Cross Hatch Cutter (Linear)MECHANICAL- DIMENSION (BASIC MEASURING INSTRUMENT, GAUGE ETC.)Permanent facility · entry 870 to 3 mmUsing VMM (Vision Measuring Machine) by Comparison Method9.56 µmPage 16
Cube Mould /Cylindrical Mould /Bar Mould (Length, Width, Height - Linear)MECHANICAL- DIMENSION (BASIC MEASURING INSTRUMENT, GAUGE ETC.)Permanent facility · entry 8850 mm X 50 mm X 50 mm to 150 mm X 150 mm X 150 mmUsing Digimatic Caliper by Comparison Method25.7 µmPage 16
Cylindrical Measuring Pin / Pin GaugeMECHANICAL- DIMENSION (BASIC MEASURING INSTRUMENT, GAUGE ETC.)Permanent facility · entry 890.5 mm to 20 mmUsing Slip Gauge Set, LVDT Probe, Comparator Stand by Comparison Method2.38 µmPage 16
Depth Vernier Caliper - L.C.: 0.01 mmMECHANICAL- DIMENSION (BASIC MEASURING INSTRUMENT, GAUGE ETC.)Permanent facility · entry 900 to 150 mmSlip Gauge Set, Caliper Checker by Comparison Method12.43 µmPage 17
Dial / Digital Thickness Gauge - L.C.: 0.01 mm and CoarserMECHANICAL- DIMENSION (BASIC MEASURING INSTRUMENT, GAUGE ETC.)Permanent facility · entry 91Up to 25 mmUsing Slip Gauge Set by Comparison Method6.74 µmPage 17
Feeler GaugeMECHANICAL- DIMENSION (BASIC MEASURING INSTRUMENT, GAUGE ETC.)Permanent facility · entry 920.03 mm to 2 mmUsing Digital Micrometer by Direct Method1.52 µmPage 17
GSM Cutter (Diameter)MECHANICAL- DIMENSION (BASIC MEASURING INSTRUMENT, GAUGE ETC.)Permanent facility · entry 93112.8 mmUsing VMM (Vision Measuring Machine) by Comparison Method9.58 µmPage 17
Measuring Scale / Steel Scale - L.C.: 0.5 mmMECHANICAL- DIMENSION (BASIC MEASURING INSTRUMENT, GAUGE ETC.)Permanent facility · entry 940 to 300 mmUsing VMM (Vision Measuring Machine) by Comparison Method290.75 µmPage 17
Measuring Steel Rule / Steel Scale / Dip Stick - L.C.: 1 mmMECHANICAL- DIMENSION (BASIC MEASURING INSTRUMENT, GAUGE ETC.)Permanent facility · entry 950 to 1000 mmScale & Tape Calibration System by Comparison Method291.28 µmPage 18
Measuring Tape/Pie Tape/Dip Tape - L.C.: 1 mmMECHANICAL- DIMENSION (BASIC MEASURING INSTRUMENT, GAUGE ETC.)Permanent facility · entry 960 to 50 mScale & Tape Calibration System by Comparison Method190 x Sqrt (L) µm, where L (L > 1 m) in mPage 18
Plain Plug Gauge / Air Plug Gauge (Cylindrical & Spherical) / Length Gauge / Flush Pin Gauge / Setting Plug GaugeMECHANICAL- DIMENSION (BASIC MEASURING INSTRUMENT, GAUGE ETC.)Permanent facility · entry 975 mm to 25 mmUsing Slip Gauge Set, Plunger Dial gauge, Comparator Stand by Comparison Method1.56 µmPage 18
Plunger Dial Gauge - L.C.: 0.001 mm and CoarserMECHANICAL- DIMENSION (BASIC MEASURING INSTRUMENT, GAUGE ETC.)Permanent facility · entry 98Up to 25 mmUsing Dial Calibration Tester by Comparison Method3.52 µmPage 18
Plunger Dial Gauge - L.C.: 0.01 mm and CoarserMECHANICAL- DIMENSION (BASIC MEASURING INSTRUMENT, GAUGE ETC.)Permanent facility · entry 99Up to 50 mmUsing Dial Calibration Tester, Slip Gauge Set, Comparator Stand by Comparison Method6.76 µmPage 18
Puppy Dial Gauge - L.C.: 0.001 mm & CoarserMECHANICAL- DIMENSION (BASIC MEASURING INSTRUMENT, GAUGE ETC.)Permanent facility · entry 100Up to 0.14 mmUsing Dial Calibration Tester by Comparison Method2.54 µmPage 19
Radius GaugeMECHANICAL- DIMENSION (BASIC MEASURING INSTRUMENT, GAUGE ETC.)Permanent facility · entry 1010.1 mm to 100 mmUsing VMM (Vision Measuring Machine) by Direct Method9.57 µmPage 19
Setting RodMECHANICAL- DIMENSION (BASIC MEASURING INSTRUMENT, GAUGE ETC.)Permanent facility · entry 10225 mm to 150 mmUsing slip gauge set, Plunger Dial Gauge, Comparator Stand by Comparison Method1.89 µmPage 19
Standard FoilMECHANICAL- DIMENSION (BASIC MEASURING INSTRUMENT, GAUGE ETC.)Permanent facility · entry 1030.05 mm to 2 mmLVDT Probe, Comparator stand by Direct Method2.35 µmPage 19
Taper Scale - L.C.: 0.1 mmMECHANICAL- DIMENSION (BASIC MEASURING INSTRUMENT, GAUGE ETC.)Permanent facility · entry 1041 mm to 15 mmUsing VMM (Vision Measuring Machine) by Comparison Method58.29 µmPage 19
Template Gauge / Jig Fixture (Angle)MECHANICAL- DIMENSION (BASIC MEASURING INSTRUMENT, GAUGE ETC.)Permanent facility · entry 1050° to 180°Using VMM (Vision Measuring Machine) by Comparison Method126.26 Seconds of ArcPage 20
Template Gauge / Jig Fixture (Linear & Radius)MECHANICAL- DIMENSION (BASIC MEASURING INSTRUMENT, GAUGE ETC.)Permanent facility · entry 1060 to 300 mmUsing VMM (Vision Measuring Machine) by Comparison Method11.49 µmPage 20
Test Sieves (Aperture Size)MECHANICAL- DIMENSION (BASIC MEASURING INSTRUMENT, GAUGE ETC.)Permanent facility · entry 107> 10 mm to 150 mmUsing Digimatic Caliper by Direct Method23.38 µmPage 20
Test Sieves (Aperture Size)MECHANICAL- DIMENSION (BASIC MEASURING INSTRUMENT, GAUGE ETC.)Permanent facility · entry 1080.032 mm to 10 mmUsing VMM (Vision Measuring Machine) by Direct Method9.55 µmPage 20
Thread Pitch Gauge (Angle)MECHANICAL- DIMENSION (BASIC MEASURING INSTRUMENT, GAUGE ETC.)Permanent facility · entry 10955° & 60°Using VMM (Vision Measuring Machine) by Direct Method126.26 Seconds of ArcPage 20
Thread Pitch Gauge (Pitch)MECHANICAL- DIMENSION (BASIC MEASURING INSTRUMENT, GAUGE ETC.)Permanent facility · entry 1100.25 mm to 6 mmUsing VMM (Vision Measuring Machine) by Direct Method9.60 µmPage 21
Ultrasonic Thickness Gauge - L.C.: 0.1 mm and CoarserMECHANICAL- DIMENSION (BASIC MEASURING INSTRUMENT, GAUGE ETC.)Permanent facility · entry 111Up to 100 mmUsing Slip Gauge Set by Comparison Method64.3 µmPage 21
Weld Gauge / Fillet Gauge - L.C.: 1°MECHANICAL- DIMENSION (BASIC MEASURING INSTRUMENT, GAUGE ETC.)Permanent facility · entry 1120° to 90°Using VMM (Vision Measuring Machine) by Comparison Method126.20 Seconds of ArcPage 21
Weld Gauge/Fillet Gauge - L.C.: 1 mmMECHANICAL- DIMENSION (BASIC MEASURING INSTRUMENT, GAUGE ETC.)Permanent facility · entry 1130 to 60 mmUsing VMM (Vision Measuring Machine) by Comparison Method9.60 µmPage 21
Wire Gauge (Linearity)MECHANICAL- DIMENSION (BASIC MEASURING INSTRUMENT, GAUGE ETC.)Permanent facility · entry 1140.19 mm to 10 mmUsing VMM (Vision Measuring Machine) by Comparison Method9.60 µmPage 21
Rubber Hardness TesterMECHANICAL- DUROMETERPermanent facility · entry 1150 to 100 Shore AUsing Dial Calibration Tester by Depth Indentation Method as per ISO 48-9: 20181.37 Shore APage 22
Rubber Hardness TesterMECHANICAL- DUROMETERPermanent facility · entry 1160 to 100 Shore DUsing Dial Calibration Tester by Depth Indentation Method as per ISO 48-9: 20181.36 Shore DPage 22
Hydraulic Medium: Analog/Digital Pressure Gauge, Pressure Transmitter, Pressure SwitchMECHANICAL- PRESSURE INDICATING DEVICESPermanent facility · entry 1170 to 700 barUsing Digital Pressure Gauge, Digital Multimeter & Hydraulic Comparator by Comparison Method based on DKD R 6-11.21 barPage 22
Pneumatic Medium: Analog/Digital Differential Gauge, Magnehelic Gauge, Pressure Gauge, Differential Pressure Transmitter, ManometerMECHANICAL- PRESSURE INDICATING DEVICESPermanent facility · entry 1180 to 2000 mbarUsing Digital Pressure Gauge, Pressure Gauge Inbuilt readout & Low Pressure Pump by Comparison Method based on DKD R 6-115.06 mbarPage 22
Pneumatic Medium: Analog/Digital Magnehelic Gauge, Pressure Gauge, Differential Pressure Transmitter, ManometerMECHANICAL- PRESSURE INDICATING DEVICESPermanent facility · entry 119(-) 200 mbar to 0Using Digital Pressure Calibrator, Pressure Gauge Inbuilt readout Low & Pressure Pump by Comparison Method based on DKD R 6-10.56 mbarPage 23
Pneumatic Medium: Analog/Digital Magnehelic Gauge, Pressure Gauge, Differential Pressure Transmitter, ManometerMECHANICAL- PRESSURE INDICATING DEVICESPermanent facility · entry 1200 to 200 mbarUsing Digital Pressure Calibrator, Pressure Gauge Inbuilt readout & Low Pressure Pump by Comparison Method based on DKD R 6-10.47 mbarPage 23
Pneumatic Medium: Analog/Digital Pressure Gauge, Pressure Transmitter, Pressure SwitchMECHANICAL- PRESSURE INDICATING DEVICESPermanent facility · entry 1210 to 34 barDigital Pressure Gauge, Pressure Gauge Inbuilt readout & Pneumatic Comparator Pump by Comparison Method based on DKD R 6-10.027 barPage 23
Pneumatic Medium: Analog/Digital Vacuum Gauge, Vacuum Transmitter, Vacuum SwitchMECHANICAL- PRESSURE INDICATING DEVICESPermanent facility · entry 122(-) 0.95 bar to 0Using Digital Pressure Gauge, Pressure Gauge Inbuilt readout & Vacuum Comparator by Comparison Method based on DKD R 6-10.006 barPage 24
Electronic Weighing Balance - (Readability: 1 g) Accuracy Class IIII and CoarserMECHANICAL- WEIGHING SCALE AND BALANCEPermanent facility · entry 1230 to 60 kgUsing E2 and F1 class Weights by Comparison Method as per OIML R-76-1: 20065 gPage 24
Electronic Weighing Balance - (Readability: 0.001 g) Accuracy Class I and CoarserMECHANICAL- WEIGHING SCALE AND BALANCEPermanent facility · entry 1240 to 220 gUsing E2 Class Weights by Comparison Method as per OIML R-76-1: 20060.008 gPage 24
Electronic Weighing Balance - (Readability: 0.1 g) Accuracy Class III and CoarserMECHANICAL- WEIGHING SCALE AND BALANCEPermanent facility · entry 1250 to 20 kgUsing E2 and F1 class Weights by Comparison Method as per OIML R-76-1: 20060.8 gPage 24
Electronic Weighing Balance - (Readability: 10 g) Accuracy Class IIII and CoarserMECHANICAL- WEIGHING SCALE AND BALANCEPermanent facility · entry 1260 to 200 kgUsing F1 class Weights by Comparison Method as per OIML R-76-1: 200618 gPage 24
Electronic Weighing Balance: (Readability: 0.01 g) Accuracy Class II and CoarserMECHANICAL- WEIGHING SCALE AND BALANCEPermanent facility · entry 1270 to 3 kgUsing E2 and F1 class Weights by Comparison Method as per OIML R-76-1: 20060.03 gPage 25
IR ThermometerTHERMAL- TEMPERATUREPermanent facility · entry 12850 °C to 500 °CUsing with Digital IR Thermometer and Black Body Source (Emissivity: 0.95) by Comparison Method4.34 °CPage 25
Liquid in Glass ThermometerTHERMAL- TEMPERATUREPermanent facility · entry 12950 °C to 250 °CUsing SSPRT Sensor with Digital Indicator & Low Temperature Liquid Bath by Comparison Method0.94 °CPage 25
RTD/Thermocouple with Temperature Indicator/Controller/ Temperature Gauge, Recorder/ Data Logger with Pt-100 Sensor, Digital Thermometer, Temperature Gauge, Temperature TransmitterTHERMAL- TEMPERATUREPermanent facility · entry 130(-) 30 °C to 100 °CUsing SSPRT Sensor with Digital Indicator, Temperature Calibrator and Low Temperature Liquid Bath by Comparison Method0.23 °CPage 25
RTD/Thermocouple with Temperature Indicator/Controller/ Temperature Gauge, Recorder/ Data Logger with Pt-100 sensor, Digital Thermometer, Temperature Gauge, Temperature TransmitterTHERMAL- TEMPERATUREPermanent facility · entry 131100 °C to 250 °CUsing SSPRT Sensor with Digital Indicator, Temperature Calibrator and Low Temperature Liquid Bath by Comparison Method0.91 °CPage 26
Thermocouples with Temperature Indicator/Controller/ Recorder/ Data Logger, Digital Thermometer, Temperature Gauge, Temperature TransmitterTHERMAL- TEMPERATUREPermanent facility · entry 132250 °C to 650 °CUsing S Type Thermocouple with Digital Indicator, Temperature Calibrator and Dry Block furnace by Comparison Method2.91 °CPage 26
Thermocouples with Temperature Indicator/Controller/ Recorder/ Data Logger, Digital Thermometer, Temperature Gauge, Temperature Transmitter with sensorTHERMAL- TEMPERATUREPermanent facility · entry 133650 °C to 1200 °CUsing S Type Thermocouple with Digital Indicator, Temperature Calibrator and Dry Block Furnace by Comparison Method3.58 °CPage 26
AC Current @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Measure)Site facility · entry 11 A to 10 AUsing 5½ Digital Multimeter by Direct Method1.28 %Page 27
AC Current @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Measure)Site facility · entry 210 mA to 100 mAUsing 5½ Digital Multimeter by Direct Method0.41 % to 0.43 %Page 27
AC Current @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Measure)Site facility · entry 3100 mA to 200 mAUsing 5½ Digital Multimeter by Direct Method0.43 % to 0.38 %Page 27
AC Current @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Measure)Site facility · entry 42 mA to 10 mAUsing 5½ Digital Multimeter by Direct Method1.31 % to 0.41 %Page 27
AC Voltage @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Measure)Site facility · entry 51 V to 10 VUsing 5½ Digital Multimeter by Direct Method0.31 %Page 28
AC Voltage @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Measure)Site facility · entry 610 V to 100 VUsing 5½ Digital Multimeter by Direct Method0.31 % to 0.30 %Page 28
AC Voltage @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Measure)Site facility · entry 7100 mV to 200 mVUsing 5½ Digital Multimeter by Direct Method0.30 % to 0.24 %Page 28
AC Voltage @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Measure)Site facility · entry 8100 V to 500 VUsing 5½ Digital Multimeter by Direct Method0.30 % to 0.28 %Page 28
AC Voltage @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Measure)Site facility · entry 920 mV to 100 mVUsing 5½ Digital Multimeter by Direct Method0.79 % to 0.30 %Page 28
AC Voltage @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Measure)Site facility · entry 10500 V to 750 VUsing 5½ Digital Multimeter by Direct Method0.28 % to 0.31 %Page 29
AC Current @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Source)Site facility · entry 110.2 mA to 1 mAUsing 5½ Multifunction Calibrator by Direct Method0.64 % to 0.26 %Page 29
AC Current @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Source)Site facility · entry 121 A to 10 AUsing 5½ Multifunction Calibrator by Direct Method0.36 % to 0.22 %Page 29
AC Current @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Source)Site facility · entry 131 mA to 20 mAUsing 5½ Multifunction Calibrator by Direct Method0.26 % to 0.27 %Page 29
AC Current @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Source)Site facility · entry 14100 mA to 200 mAUsing 5½ Multifunction Calibrator by Direct Method0.20 % to 0.027 %Page 29
AC Current @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Source)Site facility · entry 1520 mA to 100 mAUsing 5½ Multifunction Calibrator by Direct Method0.27 % to 0.20 %Page 30
AC Current @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Source)Site facility · entry 16200 mA to 2000 mAUsing 5½ Multifunction Calibrator by Direct Method0.027 % to 0.42 %Page 30
AC Current @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Source)Site facility · entry 1710 A to 700 AUsing 5½ Multifunction Calibrator with Current Coil by Direct Method2.05 % to 0.97 %Page 30
AC Voltage @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Source)Site facility · entry 181 V to 10 VUsing 5½ Multifunction Calibrator by Direct Method0.97 % to 0.44 %Page 30
AC Voltage @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Source)Site facility · entry 1910 mV to 100 mVUsing 5½ Multifunction Calibrator by Direct Method0.24 % to 0.19 %Page 30
AC Voltage @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Source)Site facility · entry 2010 V to 20 VUsing 5½ Multifunction Calibrator by Direct Method0.44 % to 0.33 %Page 30
AC Voltage @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Source)Site facility · entry 21100 mV to 200 mVUsing 5½ Multifunction Calibrator by Direct Method0.19 % to 0.21 %Page 31
AC Voltage @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Source)Site facility · entry 22100 V to 200 VUsing 5½ Multifunction Calibrator by Direct Method0.20 % to 0.19 %Page 31
AC Voltage @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Source)Site facility · entry 2320 V to 100 VUsing 5½ Multifunction Calibrator by Direct Method0.33 % to 0.20 %Page 31
AC Voltage @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Source)Site facility · entry 24200 V to 500 VUsing 5½ Multifunction Calibrator by Direct Method0.19 % to 0.21 %Page 31
AC Voltage @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Source)Site facility · entry 25500 V to 900 VUsing 5½ Multifunction Calibrator by Direct Method0.21 % to 0.18 %Page 31
AC Voltage @ 50 HzELECTRO- TECHNICAL- Alternating Current (< 1 GHz) (Source)Site facility · entry 26900 V to 1000 VUsing 5½ Multifunction Calibrator by Direct Method0.18 % to 0.20 %Page 31
DC CurrentELECTRO- TECHNICAL- DIRECT CURRENT (Measure)Site facility · entry 271 A to 10 AUsing 5½ Digital Multimeter by Direct Method0.13 % to 0.28 %Page 32
DC CurrentELECTRO- TECHNICAL- DIRECT CURRENT (Measure)Site facility · entry 281 mA to 10 mAUsing 5½ Digital Multimeter by Direct Method0.62 % to 0.084 %Page 32
DC CurrentELECTRO- TECHNICAL- DIRECT CURRENT (Measure)Site facility · entry 2910 mA to 100 mAUsing 5½ Digital Multimeter by Direct Method0.084 % to 0.078 %Page 32
DC CurrentELECTRO- TECHNICAL- DIRECT CURRENT (Measure)Site facility · entry 30100 mA to 200 mAUsing 5½ Digital Multimeter by Direct Method0.078 % to 0.076 %Page 32
DC CurrentELECTRO- TECHNICAL- DIRECT CURRENT (Measure)Site facility · entry 31200 mA to 1 AUsing 5½ Digital Multimeter by Direct Method0.076 % to 0.13 %Page 32
DC VoltageELECTRO- TECHNICAL- DIRECT CURRENT (Measure)Site facility · entry 321 mV to 200 mVUsing 5½ Digital Multimeter by Direct Method0.71 % to 0.017 %Page 32
DC VoltageELECTRO- TECHNICAL- DIRECT CURRENT (Measure)Site facility · entry 331 V to 1000 VUsing 5½ Digital Multimeter by Direct Method0.061 % to 0.02 %Page 33
DC VoltageELECTRO- TECHNICAL- DIRECT CURRENT (Measure)Site facility · entry 34200 mV to 1 VUsing 5½ Digital Multimeter by Direct Method0.017 % to 0.061 %Page 33
Resistance (2 Wire)ELECTRO- TECHNICAL- DIRECT CURRENT (Measure)Site facility · entry 351 kohm to 10 kohmUsing 5½ Digital Multimeter by Direct Method0.030 % to 0.026 %Page 33
Resistance (2 Wire)ELECTRO- TECHNICAL- DIRECT CURRENT (Measure)Site facility · entry 361 Mohm to 10 MohmUsing 5½ Digital Multimeter by Direct Method0.13 % to 0.27 %Page 33
Resistance (2 Wire)ELECTRO- TECHNICAL- DIRECT CURRENT (Measure)Site facility · entry 3710 kohm to 100 kohmUsing 5½ Digital Multimeter by Direct Method0.026 % to 0.027 %Page 33
Resistance (2 Wire)ELECTRO- TECHNICAL- DIRECT CURRENT (Measure)Site facility · entry 3810 Mohm to 90 MohmUsing 5½ Digital Multimeter by Direct Method0.27 % to 4.72 %Page 33
Resistance (2 Wire)ELECTRO- TECHNICAL- DIRECT CURRENT (Measure)Site facility · entry 39100 kohm to 1 MohmUsing 5½ Digital Multimeter by Direct Method0.027 % to 0.13 %Page 34
Resistance (4 Wire)ELECTRO- TECHNICAL- DIRECT CURRENT (Measure)Site facility · entry 4020 ohm to 180 ohmUsing 5½ Digital Multimeter by Direct Method0.700 % to 0.023 %Page 34
DC CurrentELECTRO- TECHNICAL- DIRECT CURRENT (Source)Site facility · entry 410.1 mA to 1 mAUsing 5½ Multifunction Calibrator by Direct Method0.7 % to 0.13 %Page 34
DC CurrentELECTRO- TECHNICAL- DIRECT CURRENT (Source)Site facility · entry 421 A to 10 AUsing 5½ Multifunction Calibrator by Direct Method0.24 % to 0.16 %Page 34
DC CurrentELECTRO- TECHNICAL- DIRECT CURRENT (Source)Site facility · entry 431 mA to 2 mAUsing 5½ Multifunction Calibrator by Direct Method0.13 %Page 34
DC CurrentELECTRO- TECHNICAL- DIRECT CURRENT (Source)Site facility · entry 4410 A to 900 AUsing 5½ Multifunction Calibrator by Direct Method1.91 % to 0.90 %Page 34
DC CurrentELECTRO- TECHNICAL- DIRECT CURRENT (Source)Site facility · entry 452 mA to 20 mAUsing 5½ Multifunction Calibrator by Direct Method0.16 % to 0.071 %Page 35
DC CurrentELECTRO- TECHNICAL- DIRECT CURRENT (Source)Site facility · entry 4620 mA to 200 mAUsing 5½ Multifunction Calibrator by Direct Method0.071 % to 0.14 %Page 35
DC CurrentELECTRO- TECHNICAL- DIRECT CURRENT (Source)Site facility · entry 47200 mA to 1 AUsing 5½ Multifunction Calibrator by Direct Method0.140 % to 0.24 %Page 35
DC VoltageELECTRO- TECHNICAL- DIRECT CURRENT (Source)Site facility · entry 480.2 V to 1000 VUsing 5½ Multifunction Calibrator by Direct Method0.18 % to 0.13 %Page 35
DC VoltageELECTRO- TECHNICAL- DIRECT CURRENT (Source)Site facility · entry 491 mV to 200 mVUsing 5½ Multifunction Calibrator By Direct Method1.40 % to 0.13 %Page 35
Resistance (2 Wire)ELECTRO- TECHNICAL- DIRECT CURRENT (Source)Site facility · entry 501 kohm to 10 kohmUsing Resistance Decade box by Direct Method0.14 % to 0.12 %Page 35
Resistance (2 Wire)ELECTRO- TECHNICAL- DIRECT CURRENT (Source)Site facility · entry 511 Mohm to 10 MohmUsing Resistance Decade box by Direct Method0.12 % to 3.52 %Page 36
Resistance (2 Wire)ELECTRO- TECHNICAL- DIRECT CURRENT (Source)Site facility · entry 5210 kohm to 100 kohmUsing Resistance Decade box by Direct Method0.12 %Page 36
Resistance (2 Wire)ELECTRO- TECHNICAL- DIRECT CURRENT (Source)Site facility · entry 5310 Mohm to 100 MohmUsing Decade Resistance Box by Direct Method3.52 % to 2.40 %Page 36
Resistance (2 Wire)ELECTRO- TECHNICAL- DIRECT CURRENT (Source)Site facility · entry 54100 kohm to 1 MohmUsing Resistance Decade box by Direct Method0.12 %Page 36
Resistance (2 Wire)ELECTRO- TECHNICAL- DIRECT CURRENT (Source)Site facility · entry 55100 ohm to 1 kohmUsing Resistance Decade box by Direct Method0.12 % to 0.14 %Page 36
Resistance (4 Wire)ELECTRO- TECHNICAL- DIRECT CURRENT (Source)Site facility · entry 5610 ohm to 100 ohmUsing Resistance Decade Box by Direct Method0.13 % to 0.12 %Page 36
B Type ThermocoupleELECTRO- TECHNICAL- TEMPERATURE SIMULATION (Source)Site facility · entry 57600 °C to 1800 °CUsing Universal calibrator by Direct Method3.45 °CPage 37
E Type ThermocoupleELECTRO- TECHNICAL- TEMPERATURE SIMULATION (Source)Site facility · entry 58(-) 200 °C to 1000 °CUsing Universal Calibrator by Direct Method2.90 °CPage 37
J Type ThermocoupleELECTRO- TECHNICAL- TEMPERATURE SIMULATION (Source)Site facility · entry 59(-) 200 °C to 1200 °CUsing Universal Calibrator by Direct Method3.08 °CPage 37
K Type ThermocoupleELECTRO- TECHNICAL- TEMPERATURE SIMULATION (Source)Site facility · entry 60(-) 200 °C to 1370 °CUsing Universal Calibrator by Direct Method3.19 °CPage 37
N Type ThermocoupleELECTRO- TECHNICAL- TEMPERATURE SIMULATION (Source)Site facility · entry 61(-) 200 °C to 1300 °CUsing Universal Calibrator by Direct Method2.24 °CPage 37
R Type ThermocoupleELECTRO- TECHNICAL- TEMPERATURE SIMULATION (Source)Site facility · entry 620 °C to 1750 °CUsing Universal Calibrator by Direct Method3.38 °CPage 37
RTD (Pt 100)ELECTRO- TECHNICAL- TEMPERATURE SIMULATION (Source)Site facility · entry 63(-) 200 °C to 800 °CUsing Universal Calibrator by Direct Method1.32 °CPage 38
S Type ThermocoupleELECTRO- TECHNICAL- TEMPERATURE SIMULATION (Source)Site facility · entry 640 °C to 1750 °CUsing Universal Calibrator by Direct Method3.54 °CPage 38
T Type ThermocoupleELECTRO- TECHNICAL- TEMPERATURE SIMULATION (Source)Site facility · entry 65(-) 200 °C to 400 °CUsing Universal Calibrator by Direct Method2.31 °CPage 38
FrequencyELECTRO- TECHNICAL- TIME & FREQUENCY (Measure)Site facility · entry 6645 Hz to 500 HzUsing 5½ Digital Multimeter by Direct Method0.48 % to 0.034 %Page 38
FrequencyELECTRO- TECHNICAL- TIME & FREQUENCY (Measure)Site facility · entry 67500 Hz to 1000 HzUsing 5½ Digital Multimeter by Direct Method0.034 % to 0.063 %Page 38
TimeELECTRO- TECHNICAL- TIME & FREQUENCY (Measure)Site facility · entry 680.1 s to 60 sUsing Time Calibration Standard by Direct Method0.01 s to 0.03 sPage 38
TimeELECTRO- TECHNICAL- TIME & FREQUENCY (Measure)Site facility · entry 693600 s to 86400 sUsing Time Calibration Standard by Direct Method0.79 s to 5.58 sPage 39
TimeELECTRO- TECHNICAL- TIME & FREQUENCY (Measure)Site facility · entry 7060 s to 3600 sUsing Time Calibration Standard by Direct Method0.03 s to 0.79 sPage 39
FrequencyELECTRO- TECHNICAL- TIME & FREQUENCY (Source)Site facility · entry 7145 Hz to 500 HzUsing 5½ Multifunction Calibrator by Direct Method0.30 % to 0.026 %Page 39
FrequencyELECTRO- TECHNICAL- TIME & FREQUENCY (Source)Site facility · entry 72500 Hz to 1000 HzUsing 5½ Multifunction Calibrator by Direct Method0.026 % to 0.02 %Page 39
Digital Tachometer (Non Contact-Type)MECHANICAL- ACCELERATION AND SPEEDSite facility · entry 73> 100 rpm to 1000 rpmUsing Digital Tachometer with RPM Source by Comparison Method4.01 % of rdgPage 39
Digital Tachometer (Non Contact-Type)MECHANICAL- ACCELERATION AND SPEEDSite facility · entry 74> 1000 rpm to 90000 rpmUsing Digital Tachometer with RPM Source by Comparison Method4.00 % of rdgPage 39
Digital Tachometer (Non Contact-Type)MECHANICAL- ACCELERATION AND SPEEDSite facility · entry 7510 rpm to 100 rpmUsing Digital Tachometer with RPM Source by Comparison Method5.6 % of rdgPage 40
Hydraulic Medium: Analog/Digital Pressure Gauge, Pressure Transmitter, Pressure SwitchMECHANICAL- PRESSURE INDICATING DEVICESSite facility · entry 760 to 700 barUsing Digital Pressure Gauge, Digital Multimeter & Hydraulic Comparator by Comparison Method based on DKD R 6-11.21 barPage 40
Pneumatic Medium: Analog/Digital Differential Gauge, Magnehelic Gauge, Pressure Gauge, Differential Pressure Transmitter, ManometerMECHANICAL- PRESSURE INDICATING DEVICESSite facility · entry 770 to 2000 mbarUsing Digital Pressure Gauge, Pressure Gauge Inbuilt readout & Low Pressure Pump by Comparison Method based on DKD R 6-115.06 mbarPage 40
Pneumatic Medium: Analog/Digital Magnehelic Gauge, Pressure Gauge, Differential Pressure Transmitter, ManometerMECHANICAL- PRESSURE INDICATING DEVICESSite facility · entry 78(-) 200 mbar to 0Using Digital Pressure Calibrator, Pressure Gauge Inbuilt readout Low & Pressure Pump by Comparison Method based on DKD R 6-10.56 mbarPage 40
Pneumatic Medium: Analog/Digital Magnehelic Gauge, Pressure Gauge, Differential Pressure Transmitter, ManometerMECHANICAL- PRESSURE INDICATING DEVICESSite facility · entry 790 to 200 mbarUsing Digital Pressure Calibrator, Pressure Gauge Inbuilt readout & Low Pressure Pump by Comparison Method based on DKD R 6-10.47 mbarPage 41
Pneumatic Medium: Analog/Digital Pressure Gauge, Pressure Transmitter, Pressure SwitchMECHANICAL- PRESSURE INDICATING DEVICESSite facility · entry 800 to 34 barDigital Pressure Gauge, Pressure Gauge Inbuilt readout & Pneumatic Comparator Pump by Comparison Method based on DKD R 6-10.027 barPage 41
Pneumatic Medium: Analog/Digital Vacuum Gauge, Vacuum Transmitter, Vacuum SwitchMECHANICAL- PRESSURE INDICATING DEVICESSite facility · entry 81(-) 0.95 bar to 0Using Digital Pressure Gauge, Pressure Gauge Inbuilt readout & Vacuum Comparator by Comparison Method based on DKD R 6-10.006 barPage 41
Electronic Weighing Balance - (Readability: 1 g) Accuracy Class IIII and CoarserMECHANICAL- WEIGHING SCALE AND BALANCESite facility · entry 820 to 60 kgUsing E2 and F1 class Weights by Comparison Method as per OIML R-76-1: 20065 gPage 41
Electronic Weighing Balance - (Readability: 0.001 g) Accuracy Class I and CoarserMECHANICAL- WEIGHING SCALE AND BALANCESite facility · entry 830 to 220 gUsing E2 Class Weights by Comparison Method as per OIML R-76-1: 20060.008 gPage 42
Electronic Weighing Balance - (Readability: 0.1 g) Accuracy Class III and CoarserMECHANICAL- WEIGHING SCALE AND BALANCESite facility · entry 840 to 20 kgUsing E2 and F1 class Weights by Comparison Method as per OIML R-76-1: 20060.8 gPage 42
Electronic Weighing Balance - (Readability: 10 g) Accuracy Class IIII and CoarserMECHANICAL- WEIGHING SCALE AND BALANCESite facility · entry 850 to 200 kgUsing F1 class Weights by Comparison Method as per OIML R-76-1: 200618 gPage 42
Electronic Weighing Balance: (Readability: 0.01 g) Accuracy Class II and CoarserMECHANICAL- WEIGHING SCALE AND BALANCESite facility · entry 860 to 3 kgUsing E2 and F1 class Weights by Comparison Method as per OIML R-76-1: 20060.03 gPage 42
Indicator with Sensor of Liquid Bath / Furnace / Oven / Dry Block Bath (Single Position)THERMAL- TEMPERATURESite facility · entry 87100 °C to 250 °CUsing SSPRT Sensor with Digital Indicator by Comparison Method0.91 °CPage 42
Indicator with Sensor of Liquid Bath / Furnace / Oven / Dry Block Bath (Single Position)THERMAL- TEMPERATURESite facility · entry 88250 °C to 400 °CUsing SSPRT Sensor with Digital Indicator Comparison Method0.93 °CPage 43
Indicator with Sensor of Liquid Bath / Furnace / Oven / Dry Block Bath (Single Position)THERMAL- TEMPERATURESite facility · entry 89400 °C to 1200 °CUsing S-Type Thermocouple with Digital Indicator by Comparison Method3.02 °CPage 43
Indicator with Sensor of Liquid Bath / Furnace / Oven / Freezer / Dry Block Bath / Cold Room Chamber (Single Position)THERMAL- TEMPERATURESite facility · entry 90(-) 30 °C to 100 °CUsing SSPRT Sensor with Digital Indicator Comparison Method0.22 °CPage 43
IR ThermometerTHERMAL- TEMPERATURESite facility · entry 9150 °C to 500 °CUsing with Digital IR Thermometer and Black Body Source (Emissivity: 0.95) by Comparison Method4.34 °CPage 43
Liquid in Glass ThermometerTHERMAL- TEMPERATURESite facility · entry 9250 °C to 250 °CUsing SSPRT Sensor with Digital Indicator & Low Temperature Liquid Bath by Comparison Method0.94 °CPage 43
RTD/Thermocouple with Temperature Indicator/Controller/ Temperature Gauge, Recorder/ Data Logger with Pt-100 sensor, Digital Thermometer, Temperature Gauge, Temperature TransmitterTHERMAL- TEMPERATURESite facility · entry 93100 °C to 250 °CUsing SSPRT Sensor with Digital Indicator, Temperature Calibrator and Low Temperature Liquid Bath by Comparison Method0.91 °CPage 44
Thermocouples with Temperature Indicator/Controller/ Recorder/ Data Logger, Digital Thermometer, Temperature Gauge, Temperature TransmitterTHERMAL- TEMPERATURESite facility · entry 94250 °C to 650 °CUsing S Type Thermocouple with Digital Indicator, Temperature Calibrator and Dry Block furnace by Comparison Method2.91 °CPage 44
Thermocouples with Temperature Indicator/Controller/ Recorder/ Data Logger, Digital Thermometer, Temperature Gauge, Temperature Transmitter with sensorTHERMAL- TEMPERATURESite facility · entry 95650 °C to 1200 °CUsing S Type Thermocouple with Digital Indicator, Temperature Calibrator and Dry Block Furnace by Comparison Method3.58 °CPage 44