How the box connects to the bridge
Identify the model, the terminal markings and any screen or guard connection used at your bench. For a small capacitance, a different lead or connection can alter the reading. If you compare results with an earlier certificate, check whether it reports the value at the terminals or the change from the all-zero setting.
Single decades and combined settings
A single switch position exercises one decade. A setting made up on several dials is also worth checking if it is the one technicians use to verify a meter. Supply the dial combination as well as the total capacitance. For an intermittent reading, identify the switch position where you see it.
Frequency and measurement mode
Record the test frequency and signal level used for the comparison. If the bridge can report either series or parallel equivalent capacitance, specify the mode from your procedure. With that information on the results sheet, a technician can reproduce the measurement after the box goes back into service.
Decade capacitance box
If a bench instruction uses several dials together, provide those combinations alongside the individual decade settings. A value checked at one frequency may need another comparison at the frequency used by the receiving bridge.
For low capacitance settings, the fixture and its cable can be a noticeable part of the observed reading. Record the open-fixture correction and cable arrangement used in the customer’s routine, particularly if the discrepancy appeared after moving benches.
- Range and configuration
- Small settings can be appreciably affected by residual terminal capacitance. Record both the smallest step and the range of combinations used.
- Manufacturers and models
- IET Labs GenRad 1419 decade capacitors have selectable nominal capacitance. Terminal arrangement and frequency are part of the specification.
- What we check
- Measure residual capacitance and the selected steps at a defined frequency. Apply the correct guard and fixture convention for the model.
- Reference equipment and accuracy
- Use a calibrated capacitance bridge or LCR reference and characterised capacitors at the working frequency. Fixture stability matters at the low end.
- When to repeat the check
- Recheck after switch repair or any change in residual capacitance with all decades at zero.
Technical references IET Labs GenRad 1419 capacitance boxes