Autocollimator calibration

An autocollimator detects small changes in the angle of a reflecting surface. The return image can move in two axes, and the instrument may show angular displacement directly or against a graduated eyepiece. The mirror and optical arrangement should be identifiable with the reading.

Send an enquiry
Autocollimator — equipment illustration
Equipment illustration
On this page

From an angle reading to a straightness survey

A machine straightness survey combines successive angle readings with the spacing between positions. A calibration of the autocollimator gives the angular results used in that calculation. It cannot provide the straightness of a machine the instrument has not measured. Include both reading axes if you use them, the angular interval of interest, the reflector and the step length and sign convention from your survey sheet.

Autocollimator and reflector

State whether one or two axes require results and how the reflector is mounted during the application. Include the working angular interval and any recorded zero offset. Calibration of the angular indication and an alignment survey using that instrument are separate requests; the quotation must identify which is being reviewed.

Specifications and service checks

A two-axis readout can show a small horizontal change while the mirror is being tilted vertically. If that matters during slide alignment, record both channels during the same tilt and note which way the instrument axes face on the machine.

Range and configuration
Taylor Hobson specifies ±2,000 arcseconds for its 142-214 Ultra Wide Range model. Specify the axis and the smaller span used to measure the actual work.
Manufacturers and models
Taylor Hobson’s Ultra and Ultra Wide Range instruments illustrate why the precise model and measuring field matter.
What we check
Compare the angular indication at known mirror inclinations on each requested axis. Include positive and negative deflections around zero.
Reference equipment and accuracy
Use a calibrated indexing table or small-angle generator, stable reflector and an alignment arrangement with known uncertainty.
When to repeat the check
Review after impact or optical adjustment. Use an independent check angle if the instrument is routinely transported between machine tools.