Capacitance diaphragm gauge calibration

A capacitance diaphragm gauge measures vacuum by sensing the deflection of a diaphragm. A chamber may have several heads covering different pressure decades, although all their readings appear on the same controller.

Send an enquiry
Capacitance diaphragm gauge — equipment illustration
Equipment illustration
On this page

Matching each pressure head to its channel

Keep each head paired with its controller channel and identify any temperature-controlled head. Specify the full-scale rating of the head and the operating pressure of the chamber. If the controller changes range automatically, the report should identify which head supplied each reading.

A controller might switch from a 1 Torr head to a 10 Torr head as chamber pressure rises. At an overlap point, the selected head can change while the number on the screen barely moves. If that region matters to your process, identify the two heads and mark the pressure where the controller hands over.

Capacitance diaphragm gauges

Give the absolute-pressure points and process connection needed in the report. Keep zeroing history and prior contamination information with the instrument record. RTS calibrates the absolute-pressure indication across the selected points, retaining the sensing head’s temperature setting and zeroing conditions.

Specifications and service checks

A coating chamber may have one heated sensor head and one unheated spare with the same nominal range. Record the actual head temperature setting and its controller. A spare that has just been powered up need not show the same zero behaviour as a head already at temperature.

Range and configuration
For example, the 627F includes 1, 10 and 100 Torr versions. A low full scale gives a different usable resolution and uncertainty from a high one.
Manufacturers and models
MKS Baratron 627F is a heated absolute capacitance manometer. Record the full-scale option and its controller.
What we check
Compare the stabilised pressure indication or analogue output against known absolute pressures. Check zero under the conditions specified for that sensor.
Reference equipment and accuracy
Use an absolute pressure transfer standard with traceable values across the requested decades. Temperature stability and zero uncertainty matter at the lower points.
When to repeat the check
Review zero history, exposure to process deposits and the effects of cleaning before selecting the next interval.

Technical references 627F Heated (45°C) Absolute Baratron® Capacitance Manometers