On this page
Field size and alignment on the imaging system
If two imaging rooms use the same model, a report marked only with the collimator type leaves the installation unclear. Give the room, equipment tag and any accessory fitted when the field was checked. Where the engineering worksheet calls for alignment, identify the field edges or marks it refers to. The service record can then distinguish an observed alignment difference from a change in the indicated field size.
Collimator alignment and field definition
Give its equipment model and the alignment or field-dimension question in the imaging QA procedure. Include the responsible medical physicist and the required report. RTS measures the defined alignment or field characteristic in the specified imaging arrangement.
Specifications and service checks
The centre of a light field can look right while one edge is displaced. On a field-alignment enquiry, label the four edges by mounting orientation and identify the equipment distance used by the imaging physicist. A single width measurement discards that information.
- Range and configuration
- The manufacturer specifies a field up to 43 × 43 cm at a source-to-image distance of 100 cm. A field dimension is only meaningful with the distance recorded.
- Manufacturers and models
- Ralco’s R 72 S is a square-field X-ray collimator.
- What we check
- A qualified medical physics team can compare the indicated light field and radiation field and document alignment at a defined distance.
- Reference equipment and accuracy
- Use a suitable field alignment test object, traceable length reference and the imaging system identified in the physics report. Geometry and positioning contribute to uncertainty.
- When to repeat the check
- Follow the facility’s radiation QA programme and review alignment after work on the tube head, light source or collimator.
Technical references Ralco product reference