On this page
The counting area and reference pattern
Identify the counting area, software version where applicable and the reference artwork or pattern used by your laboratory. Keeping the reference and reported count together makes later differences easier to investigate after a display, camera or software change.
For an automated counting instrument, a reference pattern with a known count is useful independent of any specimen. Give the pattern identifier, the assigned count and whether the operator reviewed or manually corrected the displayed count. That keeps the comparison about the instrument and its counting record.
Colony counter counting and image records
State the required comparison and the reference image or control material used by the laboratory. A dimensional scale check and validation of a counting algorithm answer different questions. Keep the software version and counting settings with the enquiry, without sending identifiable clinical samples or records.
Specifications and service checks
A tally switch that registers twice can produce a wrong total even on a clearly marked reference pattern. Have the technician compare the shown count against a known sequence of marks and identify whether the fitted pen or an external switch was used for the count.
- Range and configuration
- A manual electronic counter records individual input events. Display capacity and magnification options are equipment details. A counting-channel check needs a known number of events.
- Manufacturers and models
- Cole-Parmer identifies the CC-200 as the successor designation for the Stuart SC6PLUS counter.
- What we check
- Compare the electronic count with a known tally using a suitable service check. Identify the display, switch and model so the result follows the actual instrument.
- Reference equipment and accuracy
- Use an independently recorded event count or suitable counter reference for the digital indication. Record the count used and any missed or repeated inputs.
- When to repeat the check
- Recheck after replacing a switch or display or when recorded totals are inconsistent. Keep the functional check with the service history.
Technical references Cole-Parmer CC-200 specifications