+91 70565 55582[email protected]Get a quoteRequest a call
RTS Calibration
Menu

Concrete mixer speed and timer checks

For a laboratory concrete mixer, the speed of the mixing assembly and the interval controlled by the timer can affect consistency between runs. Bowl capacity on the nameplate identifies the machine but does not specify either measurement.

Send an enquiry
Concrete mixer — equipment illustration
Equipment illustration
On this page

Bowl speed and the mixing interval

Identify the drive and bowl configuration, the speed or gear setting and whether the timing interval includes start-up. If the mixing procedure uses a separate stopwatch or weighing scale, list those assets independently. The equipment record can then follow each result back to the appropriate device.

A controller may count mixing time from the push-button press while your laboratory records time from the first drum movement. If a programme has a separate discharge period, identify that interval too. Mark the start and finish events on the worksheet so a repeat time check uses the same interval.

Concrete mixer preparation functions

The record identifies laboratory or production use, the normal batch and bowl arrangement, and each function measured. Mixing effectiveness is evaluated against its separate preparation or process criteria, alongside the equipment measurements required by the method.

Specifications and service checks

Two bowls may fit the same mixer shaft. When a lab changes bowl or paddle, keep the fitted pair's identifiers with any speed and clearance checks. Otherwise a repeat job can describe different mixing geometry.

Range and configuration
Matest lists a 56 L laboratory mixer. Pan capacity and usable batch size are different entries. For metrology, identify each selected speed and any timer used to control mixing.
Manufacturers and models
Matest C162 is a 56 L laboratory pan mixer. Humboldt lists a utility mixer with a steel drum.
What we check
Check rotational speed and timer indication. Where blade clearance is part of the applicable method, measure it with the designated bowl and blades.
Reference equipment and accuracy
Use a tachometer, timer and calibrated clearance gauges. Their ranges should cover the lowest mixer speed and the tightest specified gap.
When to repeat the check
Recheck clearance after fitting a new blade or bowl and speed after work on the drive.

Technical references Matest product reference · Humboldt product reference

Enquire about Concrete mixer

Service enquiry