How channels appear in the exported file
Export a short example with the column headings intact. It shows which units and channel names appear in the saved data. If timing matters, identify the logging interval as a separate requirement from the measured input values.
For a logger recording pressure through a transmitter and temperature through a separate probe, a useful order distinguishes the electrical current input from the temperature channel. To include the conversion to pressure, identify the pressure span and state whether the transmitter itself is part of the measurement.
Specifications and service checks
Two exports from the same logger can have different timestamps after daylight-saving or time-zone settings change. Record a known clock event next to the logging interval and the time zone in the export. This lets a reviewer align the temperature values with the right plant event.
- Range and configuration
- Testo gives −30 to +70 °C for the single-channel 174 T. A logger fitted with an external sensor needs the range of that sensor and its channel settings recorded separately.
- Manufacturers and models
- Testo’s 174 T and Fluke’s 54 II logging thermometer use different sensor and recording arrangements.
- What we check
- Compare the exported readings with a reference at stable temperatures. For external probes, mark the channel and probe serial number so the comparison can be repeated.
- Reference equipment and accuracy
- Use a stable temperature source and calibrated thermometer with adequate uncertainty at each point. Timestamp accuracy requires a separate time comparison if specified.
- When to repeat the check
- Check after a new probe or a suspected excursion beyond its rating. Previous drift and the allowable error in the stored records guide the next recall.