On this page
Jaw position at the inspection size
Which jaws or other measuring feature do your inspectors use? Include the dial graduation and the working dimension, particularly when it is close to a drawing limit. We can compare the caliper there and at other dial positions. Tell us if the bezel was reset with the jaws closed.
A dial with a worn rack tooth can misread near a particular opening. If the issue appears repeatedly on one component size, give the approximate opening and pointer position. That directs attention to the part of the rack that passes through the pinion there.
Specifications and service checks
With a scale reading whole millimetres and a dial reading hundredths, 18 mm on the beam plus 0.36 mm on the dial gives 18.36 mm. A shifted bezel affects the second part of that reading even if the beam is correctly read.
- Range and configuration
- Mitutoyo’s 505-730 dial caliper covers 0–150 mm in 0.02 mm graduations. Include points across different turns of the pointer when the full travel is in use.
- Manufacturers and models
- Mitutoyo Series 530 uses a vernier. The 505 is a dial family and 500-series units use a digital readout.
- What we check
- Check outside readings along the travel and at different jaw positions. Use separate comparisons for inside jaws, depth rod and step faces wherever those functions are used.
- Reference equipment and accuracy
- Calibrated gauge blocks provide outside lengths. Rings, step standards or a caliper checker supply geometries the blocks alone cannot reproduce. Match uncertainty to the required tolerance.
- When to repeat the check
- Close the jaws against light to check for a gap and verify zero before use. A drop, stiff slide or chipped tip warrants a check before the next inspection.
Technical references Vernier Scale Caliper - Mitutoyo · Mitutoyo Caliper reference · Mitutoyo Caliper specifications