On this page
Measurements at the jaw tips and near the beam
State the display unit and the jaw or rod used for each inspection point. A thin shaft near the jaw tips and a block placed near the beam do not contact the same part of the faces. Mark the measuring position when it matters to the component.
A caliper used on a 24 mm shaft and in a 6 mm recess has two jobs. Put both on the order: outside jaws at 24 mm, depth rod at 6 mm. The same display serves both, but the two measurements use different contact faces.
Specifications and service checks
A spacer drawing gives 50.00 ±0.05 mm. Its dimensional limits are 49.95 and 50.05 mm. A workshop reading of 50.04 mm is close to the upper limit, so the caliper’s display increment alone cannot settle a borderline inspection decision.
| Workshop measurement | What to include in the job |
|---|---|
| Width of a finished spacer | Outside jaws and the widths shown on the drawing. |
| Inside diameter of a recess | Inside jaws and the recess diameter. Say if access forces the inspector to use only the tips. |
| Depth of a blind slot | Depth rod and working depth. Include the base attachment if the inspector uses one. |
- Range and configuration
- Mitutoyo 500-719-20 is a 0–150 mm digital caliper fitted with a round depth rod. Its outside, inside and depth functions need points that reflect the parts being inspected.
- Manufacturers and models
- Mitutoyo Series 530 uses a vernier. The 505 is a dial family and 500-series units use a digital readout.
- What we check
- Check outside readings along the travel and at different jaw positions. Use separate comparisons for inside jaws, depth rod and step faces wherever those functions are used.
- Reference equipment and accuracy
- Calibrated gauge blocks provide outside lengths. Rings, step standards or a caliper checker supply geometries the blocks alone cannot reproduce. Match uncertainty to the required tolerance.
- When to repeat the check
- Close the jaws against light to check for a gap and verify zero before use. A drop, stiff slide or chipped tip warrants a check before the next inspection.
Technical references Vernier Scale Caliper - Mitutoyo · Mitutoyo Caliper reference · Mitutoyo Caliper specifications