Live readings and stored extremes
For an instrument with maximum and minimum memory, identify whether those stored values also matter. A comparison of the live display does not establish when an extreme occurred or whether it was retained after a power interruption.
If two units read 49% and 51% at different ends of a store, their difference alone does not reveal which one is in error. Simultaneous readings against a common reference at one location answer the instrument question. A room survey would address the separate question of a spatial difference.
Specifications and service checks
A wall unit stores a maximum RH value as well as the current reading. A maximum carried over from yesterday can look like a persistent disagreement with a handheld reference. Identify which display was recorded and the time the memory was last cleared.
- Range and configuration
- Relative humidity is reported in %RH at a stated temperature. A Vaisala HMP110 probe spans 0–100 %RH, with accuracy specified in bands and under stated conditions.
- Manufacturers and models
- The Vaisala HM41 pairs an HM40 display with an HMP113 probe fitted directly to it. If the probe has been replaced since the last calibration, record the new probe number alongside the display serial number.
- What we check
- Compare the humidity shown on the display with the reference value after both have stabilised. If the meter also calculates dew point, identify that output separately. Its value depends on the temperature reading as well as the humidity sensor.
- Reference equipment and accuracy
- Use a calibrated reference hygrometer or humidity generator with a suitable uncertainty at the actual temperature. Display resolution alone does not establish accuracy.
- When to repeat the check
- Review drift at the humidity levels used. Frequent condensation, contamination or sensor replacement can justify a shorter interval.