On this page
Matching pulses to shaft revolutions
Identify the pickup, pulses-per-revolution setting and required speed points. State whether you need calibration of the display input alone or the complete sensor and readout. Include any low-speed range where the display takes longer to update.
A readout set for one pulse per revolution would report twice the actual speed if the pickup supplies two pulses each turn. Checking a known pulse frequency together with the configured factor can distinguish a setup error from an incorrect timebase.
Specifications and service checks
For a variable-speed machine, the display averaging period affects how quickly a tachometer follows a speed change. If it agrees at steady speed but lags during a ramp, give the ramp duration and display mode. A static speed comparison addresses the first observation.
- Range and configuration
- One pulse per revolution at 60 Hz represents 3,600 rpm. Extra reflective marks or a different pulses-per-revolution setting change the conversion.
- Manufacturers and models
- Fluke 930 and 931 have an optical mode spanning 1–99,999 rpm. The figure applies to optical speed measurement under the manufacturer's conditions.
- What we check
- Compare indication with a characterised rotating target or controlled optical pulse rate over the required speed band. A contact function needs its own mechanical comparison.
- Reference equipment and accuracy
- For a combined contact and optical model use the correct speed reference for each mode. A light-pulse source can verify the electronic optical channel while a reference shaft includes the mechanical adapter.
- When to repeat the check
- Check target detection between calibrations and review after sensor damage or persistent unstable readings.
Technical references Fluke 930 and 931 tachometer manual