Disintegration tester qualification

A tablet disintegration tester has a moving basket assembly, a timed motion and a temperature-controlled liquid environment on units fitted with a bath. The quantities relevant to the equipment check include basket travel, movement rate, time and temperature.

Send an enquiry
Disintegration tester — equipment illustration
Equipment illustration
On this page

Basket travel, timing and bath temperature

On a two-basket unit, does the timer stop both baskets together or can the operator time them separately? Tell us which basket and timer the method uses. The motion check concerns stroke length and cycles per minute. The bath and, when required, the liquid in the working vessel are possible temperature measurement locations. Mark the one your procedure specifies.

Disintegration tester movement and temperature

Identify the specified movement, timing, dimensions and temperature conditions requiring checks. If automatic endpoint detection is included, describe that as a separate functional requirement. RTS measures the specified basket movement, timing and temperature with the complete apparatus configuration recorded.

Specifications and service checks

The basket timer can be correct even if a station travels a different distance on each stroke. List elapsed time, stroke rate and travel as separate checks and identify any particular station of concern. A comparison of the timer alone cannot explain a mechanical change at that position.

Range and configuration
ERWEKA’s ZT 120 Light has versions with one or two test stations. Basket travel, stroke rate, time and bath temperature are separate equipment measurements.
Manufacturers and models
ERWEKA’s ZT 120 Light series uses motor-driven stations and an integral heated bath.
What we check
Compare motion, timing and temperature at the stations named on the order. Report the individual results against the acceptance limits specified for the job.
Reference equipment and accuracy
Use suitable calibrated length, time and temperature references. A stable bath display gives no independent result for the basket stroke or elapsed time.
When to repeat the check
Review motion after drive repairs and temperature after heater or probe service. Use the applicable laboratory schedule and prior results to set the recurring checks.

Technical references ERWEKA ZT 120 Light