On this page
A single delivery or a complete series
For repeat dispensing, state whether the requirement concerns each individual delivery or the total delivered over a series. Record the attachment, volume setting and delivery sequence in the results. A total alone can hide differences between individual dispenses.
Ten dispenses collected in one vessel give the total volume delivered. Ten individually measured dispenses can show whether the first delivery differs from the later ones. Tell RTS which information you need before ordering, especially when only the first delivery after a pause is used in your process.
Specifications and service checks
Sometimes the first delivery after a pause is smaller than later ones. In that case, identify the outlet and the length of the usual pause and ask about comparing the first delivery with subsequent ones. A single total from several strokes hides the pattern.
- Range and configuration
- A Dispensette S can carry a fixed 10 mL setting, or an adjustable scale such as 1–10 mL. Specify the dose on the actual dispenser and whether the outlet assembly travels with it.
- Manufacturers and models
- BRAND’s Dispensette S family includes fixed-volume dispensers. Its Titrette is a different device intended for titration.
- What we check
- Compare repeated deliveries at the marked working settings, identifying the fitted outlet and valves. Report both mean dose and spread.
- Reference equipment and accuracy
- Choose a calibrated balance appropriate to the collected mass. Temperature and any loss from evaporation affect conversion to the reported volume.
- When to repeat the check
- A change of piston, valve, outlet or dose setting can justify an earlier check. Use delivery history to set the routine interval.
Technical references BRAND Dispensette S manual · BRAND Titrette manual · ISO 8655-6 gravimetric volume determination · BRAND Dispensette S volume settings