Probe identity and compensation settings
Look at the unit on the last worksheet. A result in mg/L is a concentration, while a reading in percent saturation also needs the conditions used in the conversion. Send the probe with the meter, identify any separate temperature sensor and note the pressure or salinity setting where the instrument uses one. That helps us investigate a mismatch in one display mode when the other looks normal.
Dissolved oxygen meter and probe
State the reported unit, temperature measurement and any pressure or salinity settings used by the method. These conditions affect interpretation of the reading. Send the reference approach and requested points for review; a meter’s electrical input check alone does not examine the complete dissolved-oxygen sensing system.
Specifications and service checks
A service log in mg/L and an instrument display in percent saturation are recording different quantities. Identify which output operators use and retain the instrument’s temperature, salinity and pressure settings where applicable. That makes comparison with the original observation possible.
- Range and configuration
- YSI specifies 0–50 mg/L for dissolved oxygen on the Pro20i. Percentage saturation and mass concentration are different display modes with different compensation requirements.
- Manufacturers and models
- YSI’s Pro20i is a portable meter with an identified probe and cable specification.
- What we check
- For a detached meter and probe, identify the required display mode, original probe and temperature channel. Compare with an appropriate assigned reference condition under the manufacturer’s instrument procedure.
- Reference equipment and accuracy
- The reference may require traceable temperature and pressure readings as well as the assigned oxygen condition. Identify the reference and corrections used on the service report.
- When to repeat the check
- Review after a membrane, sensor cap or probe replacement as applicable. Use routine check readings and the maker’s maintenance guidance for periodic recall.
Technical references YSI Pro20i specifications