Distillation unit qualification

A laboratory distillation unit may include separate heating, timing and temperature-indication functions. Define which installed quantity requires checking rather than requesting a single calibration value for the assembly. A delivery volume or output-rate requirement, where specified, is another distinct measurement.

Send an enquiry
Distillation unit — equipment illustration
Equipment illustration
On this page

Temperature, collected volume and elapsed time

Identify the receiving vessel, measurement location and approved equipment procedure. Retain elapsed time and volume together for any rate result. A calibration of the temperature readout remains traceable to that channel even if other parts of the unit are replaced.

When the acceptance sheet specifies output per hour, it matters whether the operator starts the clock at power-on or after the first collection. Send the start event in your procedure with the receiving-vessel identification. A temperature reading at the outlet is a separate observation with its own sensor position.

Distillation unit measurements

List each installed indicator and relevant vessel. A temperature comparison alone does not establish distillate purity or the yield of the full process. Supply the equipment arrangement and protocol so the qualification record identifies the operating condition assessed.

Specifications and service checks

A replacement thermometer can move the sensing tip even when the mounting fitting stays in place. Sketch the tip location and identify the displayed channel if distillate temperature is the value in question. A bath-temperature check answers a separate question.

Range and configuration
BÜCHI lists 20–280 rpm for the Rotavapor R-100. A distillation unit may instead have separate indications for time, temperature or delivered volume. Identify the quantity on the actual assembly.
Manufacturers and models
BÜCHI offers Kjel Line steam-distillation units. Identify the specific unit and measuring module where the service order concerns a timer, temperature readout or dosing channel.
What we check
Compare the indicated channel with an independent reference. Specify the temperature measuring position or the particular drive being checked, with the fitted accessories identified.
Reference equipment and accuracy
Use a calibrated thermometer for a thermal channel, a suitable speed reference for rotation, or a calibrated time reference for the timer. Choose each for the working span.
When to repeat the check
Review affected readings after a heater, drive or sensor change. A spare bath or drive can have its own service history.

Technical references BÜCHI Kjel Line instruments · BÜCHI Rotavapor R-100 manual