ELISA reader testing

A microplate reader used for absorbance measurements reports a value for each well at a selected wavelength. Photometric response, wavelength selection and consistency across well positions are separate aspects of the instrument's performance.

Send an enquiry
ELISA reader — equipment illustration
Equipment illustration
On this page

Checking absorbance with a reference plate

A single reading in the centre of a plate tells you little about the corner wells. If the concern is position-dependent, mark the wells on a plate diagram and identify the read wavelength and any reference wavelength. With a certified check plate, the assigned value belongs to a particular position, wavelength and orientation. Those details let us compare the reader’s result with the right value.

ELISA reader optical performance

Specify the optical accuracy, repeatability or other performance records in the instrument protocol. A suitable reference plate or material must match those requirements. Checking reader performance is separate from validating an assay, its reagents or the interpretation of a patient result.

Specifications and service checks

A replacement filter may be fitted in a position previously assigned to a different wavelength in software. Identify the filter wheel position, its wavelength marking and the software selection together. This gives the technician three things to reconcile when a reference plate reading shifts.

Range and configuration
A plate reader has optical channels identified by wavelength or filter. A washer has a manifold and plate-format designation. Agilent lists 96- and 384-well formats for the BioTek 405 TS.
Manufacturers and models
Agilent BioTek’s ELx808 is an absorbance reader and the 405 TS is a microplate washer.
What we check
For the optical reader, identify the wavelength filters and the assigned optical reference used on the service certificate. Record the instrument serial number and which optical channels the comparison covers.
Reference equipment and accuracy
For an optical indication, use a compatible certified optical reference with stated wavelength and uncertainty. A washer service record should identify any independently calibrated measuring channel.
When to repeat the check
Review after reader filter changes or replacement of a measuring module. Retain the reference identifier and date with the equipment record.

Technical references Agilent BioTek optical reader accessories · Agilent BioTek 405 TS specifications