Set type, delivered volume and elapsed time
For a fleet inspection, list the pump model and compatible giving set together. Identify whether the record calls for delivered volume over a period, run time to a programmed volume, or operation of an alarm. Each is a different result for the same pump.
For example, a ward may send two identical pumps and only one intermittent feeding set. Label the set used on each test and the time over which delivery was collected. A later reviewer can then see whether the pumps were compared under the same conditions.
Specifications and service checks
For a bench delivery comparison, record the pump's displayed volume and the independently collected quantity over the same interval. Identify the giving set used for the run, especially if procurement has changed the set supplied with this model.
- Range and configuration
- The Moog Infinity accepts settings from 0.1 to 600 mL/h. A bench verification specifies selected rates, test duration and the administration set used.
- Manufacturers and models
- Moog Medical’s Infinity is an enteral feeding pump. The manufacturer specifies compatible feeding sets for its delivery mechanism.
- What we check
- Measure delivered quantity over a timed interval using the manufacturer’s engineering procedure. Record set identity and fluid conditions when reporting the flow error.
- Reference equipment and accuracy
- Use a calibrated balance and timebase or a compatible flow analyser with adequate low-flow uncertainty. Where mass is converted to volume, account for fluid density.
- When to repeat the check
- Use the manufacturer’s recall instructions and review following drive or door mechanism work. Record any changed set type before comparing results across visits.
Technical references Moog Medical product reference