Force-balanced piston gauge calibration

A force-balanced piston gauge balances the force exerted by pressure on a piston against a measured force. Depending on the design, an electronic system measures that force in place of the conventional stack of weights. The piston and force-measuring assembly identify the pressure reference.

Send an enquiry
Force-balanced piston gauge — equipment illustration
Equipment illustration
On this page

Identifying the piston and force measurement

A nameplate photograph and manufacturer’s description identify the principle in your instrument. Include the piston or module identifier, working gas, pressure reference and displayed units. State whether your results are needed for the complete pressure indication or for a removable component.

Where the earlier paperwork mentions only a piston base number, check whether a force-measuring module has since been replaced. Record both identities and the measured configuration. A result attached only to the base may not distinguish the two assemblies.

Specifications and service checks

For a force-balance piston system displaying pressure electronically, an assignment may concern the complete instrument or the pressure-to-force conversion assembly. Identify the component that was serviced. A replacement force sensor and a change to the piston require different follow-up measurements.

Range and configuration
The FPG8601 specifies an overall 0–15 kPa range. Identify the reference mode and installed configuration when choosing points.
Manufacturers and models
Fluke Calibration FPG8601 is a force-balanced piston gauge for low differential and absolute pressures.
What we check
Verify the generated low pressures against an independent reference. A check of the internal force readout alone does not establish the pressure generated at the test port.
Reference equipment and accuracy
A calibrated low-pressure transfer standard or independent pressure balance is required. The reference level, zero stability and assembly temperature affect uncertainty.
When to repeat the check
Use zero history and comparison data to review the interval, particularly after work on the piston or force-measuring system.

Technical references Fluke FPG8601 pressure calibration system

Enquire about Force-balanced piston gauge