Temperature results by heating zone
If several zones have separate controllers, label the zone and input channel beside every reading. For a survey, identify the charge layout and the points used to judge uniformity. This preserves the connection between an individual trace and its physical location.
For a three-zone furnace, a good result from the middle controller does not describe the two end zones. An order for checks on all three indications and an order for a survey of the loaded work zone are different purchases. State which is needed for the production record.
Specifications and service checks
A controller sensor near a furnace wall and a separate probe at the centre can show different temperatures during warm-up. Decide whether the process record uses the controller or the central working zone. Identify the agreed comparison position and the period after stabilisation.
- Range and configuration
- Carbolite Gero identifies the CWF 13/36 as a 1,300 °C chamber furnace. Separate the model’s maximum rating from its normal working temperatures and the dimensions of its useful zone.
- Manufacturers and models
- Carbolite Gero CWF laboratory furnaces and Nabertherm W-series industrial furnaces illustrate different chamber sizes and arrangements.
- What we check
- Compare independent temperature measurements with the furnace indication at the specified positions. For a zone survey, record the load arrangement and sensor locations.
- Reference equipment and accuracy
- Use calibrated thermocouples and a readout covering the requested temperatures. At high temperatures sensor drift, radiation and immersion influence uncertainty.
- When to repeat the check
- Review after an element replacement, controller change or deterioration of the lining. A reference check at one position can flag a change between surveys.
Technical references Carbolite Gero CWF chamber furnaces · Nabertherm industrial furnace catalogue