Furnace calibration

A furnace can be checked at the controller, at a reference position in the chamber or across the working zone. These are three different jobs. Start with the temperature range and the part of the furnace actually occupied during a run.

Send an enquiry
Furnace — equipment illustration
Equipment illustration
On this page

Temperature results by heating zone

If several zones have separate controllers, label the zone and input channel beside every reading. For a survey, identify the charge layout and the points used to judge uniformity. This preserves the connection between an individual trace and its physical location.

For a three-zone furnace, a good result from the middle controller does not describe the two end zones. An order for checks on all three indications and an order for a survey of the loaded work zone are different purchases. State which is needed for the production record.

Specifications and service checks

A controller sensor near a furnace wall and a separate probe at the centre can show different temperatures during warm-up. Decide whether the process record uses the controller or the central working zone. Identify the agreed comparison position and the period after stabilisation.

Range and configuration
Carbolite Gero identifies the CWF 13/36 as a 1,300 °C chamber furnace. Separate the model’s maximum rating from its normal working temperatures and the dimensions of its useful zone.
Manufacturers and models
Carbolite Gero CWF laboratory furnaces and Nabertherm W-series industrial furnaces illustrate different chamber sizes and arrangements.
What we check
Compare independent temperature measurements with the furnace indication at the specified positions. For a zone survey, record the load arrangement and sensor locations.
Reference equipment and accuracy
Use calibrated thermocouples and a readout covering the requested temperatures. At high temperatures sensor drift, radiation and immersion influence uncertainty.
When to repeat the check
Review after an element replacement, controller change or deterioration of the lining. A reference check at one position can flag a change between surveys.

Technical references Carbolite Gero CWF chamber furnaces · Nabertherm industrial furnace catalogue