Repeat surveys after an element or sensor change
If heating elements or the control sensor have recently been changed, note the date against the last survey. A subsequent measurement can then be compared with the same zone arrangement and load condition.
If a survey from before a rebuild shows distances from the old door opening, establish the corresponding reference point on the rebuilt furnace. A small change in the apparent position of a hot zone can otherwise be confused with a change in the coordinate system.
Specifications and service checks
A furnace used near the upper end of its working range can have a shorter uniform zone than the space visible through the door. Mark the length actually occupied by the work and the locations of any independent probes. The comparison can then cover the relevant part of the chamber.
- Range and configuration
- Carbolite Gero identifies the CWF 13/36 as a 1,300 °C chamber furnace. Separate the model’s maximum rating from its normal working temperatures and the dimensions of its useful zone.
- Manufacturers and models
- Carbolite Gero CWF laboratory furnaces and Nabertherm W-series industrial furnaces illustrate different chamber sizes and arrangements.
- What we check
- Compare independent temperature measurements with the furnace indication at the specified positions. For a zone survey, record the load arrangement and sensor locations.
- Reference equipment and accuracy
- A calibrated thermocouple and readout must cover the highest requested point. Review the reference probe’s exposure history and drift uncertainty before using it near its temperature limit.
- When to repeat the check
- Review after an element replacement, controller change or deterioration of the lining. A reference check at one position can flag a change between surveys.
Technical references Carbolite Gero CWF chamber furnaces · Nabertherm industrial furnace catalogue