When the holding time begins
For a timed hold, identify whether the timer starts at switch-on or once a temperature condition is met. Keeping the start event with the traces makes it possible to distinguish warm-up time from time in the accepted band.
A paper printout may show only the set temperature and the words ‘cycle complete’. Ask whether your quality team also needs the actual temperature record from the occupied shelf. On a repeat survey, identify the tray pattern, fan setting where adjustable and the position of the independent probe. The previous results then have a useful point of comparison.
Specifications and service checks
If a fan-speed switch has two positions, note which one was selected during the original survey. The same air temperature shown by the controller can accompany different temperature distributions on the shelves. Keeping fan setting and tray layout with the results makes repeat surveys easier to compare.
- Range and configuration
- Memmert’s UF110m is specified from 10 °C above ambient to +300 °C. Despatch lists +204 °C as the upper temperature for its LBB forced-convection oven. Model and airflow design matter.
- Manufacturers and models
- Memmert UF and Despatch LBB are examples of forced-convection ovens for laboratory or production work.
- What we check
- Compare the display with temperatures measured at agreed shelf positions. Record the load, fan setting and useful volume so a repeat job can reproduce the arrangement.
- Reference equipment and accuracy
- A calibrated multipoint thermometer or logger measures differences across the oven. Include sensor uncertainty and temporal variation when assessing the specified temperature tolerance.
- When to repeat the check
- A change to the circulation fan, baffles or door seal can alter the hot-air distribution even when the controller remains accurate. Review the workspace survey after those changes.
Technical references Memmert UF110m specifications · Despatch LBB forced-convection ovens