Hot plate calibration

The surface of a hot plate and the sensing point used by its controller need not be at the same temperature. Give the normal contact area and explain whether the work concerns the bare plate, a fitted block or an external control probe.

Send an enquiry
Hot plate — equipment illustration
Equipment illustration
On this page

Where to measure the surface temperature

For an indication check, choose the position where you use the plate. If you need to know how far the temperature varies across the working surface, mark several positions from a fixed edge. On plates with multiple heating zones, label each position with its zone.

Hot plates and surface temperature

Identify the requested measurement location, contact arrangement and load condition. Include any external probe used by the operator. RTS measures surface or load temperature at the selected positions; the setpoint and the measured temperature are recorded separately.

Specifications and service checks

A technician regulating a liquid from an immersed external probe is measuring a different quantity from someone using the plate’s surface setting. Specify whether the requested comparison concerns the plate centre, another surface position or the removable liquid probe, and identify the vessel if it forms part of normal use.

Range and configuration
The IKA C-MAG HP 7 specifies a heating range of +50 to +500 °C. Surface, vessel and liquid temperatures are separate measurements.
Manufacturers and models
IKA offers the C-MAG HP 7 hot plate and the C-MAG HS 7 heating stirrer.
What we check
Compare the defined surface position or the attached external sensor with a reference. Identify the vessel or test contact used whenever the result depends on it.
Reference equipment and accuracy
Choose a calibrated surface probe or contact assembly suitable for the plate. Contact resistance and position are important uncertainty terms.
When to repeat the check
A damaged plate surface or a changed control sensor warrants a check. Use results at the normal working area to review the period between visits.

Technical references IKA C-MAG HP 7 specifications · IKA C-MAG HS 7 specifications