Humidity data logger calibration

Humidity histories often appear beside temperature in one file. At each comparison point, retain the reference temperature with the relative humidity reading. The same percentage measured at two different temperatures does not describe the same moisture condition.

Send an enquiry
Humidity data logger — equipment illustration
Equipment illustration
On this page

Probe identity and the logged readings

For an external probe, keep the logger and probe identifiers together. If the evidence used by quality staff is a downloaded file, ask for checks against the saved readings and include the interval and averaging settings shown in that file.

When a logger stores a ten-minute average but another records one sample every ten minutes, their records can show different peaks during a short room disturbance. Supply the instrument settings with the file so differences caused by logging behaviour are not mistaken for calibration error.

Specifications and service checks

A logger set to save a ten-minute average can miss a brief rise visible on its live display. Check the saved interval and averaging mode before comparing the file with spot observations. Include one exported run that shows the headings and units used in your quality records.

Range and configuration
The Testo 174 H measures 0–100 %RH and −20 to +70 °C. Its humidity accuracy specification applies over a narrower band than the full display scale.
Manufacturers and models
Vaisala’s RFL100 accepts different probe types. Testo’s 174 H records temperature and humidity locally.
What we check
Compare the measured channel with a suitable environmental reference and read back the stored result. Check the remote display uses the same units and channel assignment.
Reference equipment and accuracy
Use a humidity reference with assigned uncertainty at both the working humidity and temperature. Log enough information to identify the probe, logger and time corresponding to the reported comparisons.
When to repeat the check
Review the interval when a replaceable probe changes. A radio outage calls for a review of missing records and is distinct from a probe calibration issue.

Technical references Vaisala RFL100 guide · Testo 174 data logger comparison