The input signal and its humidity scale
When requesting an indicator-only comparison, give the input range and its corresponding humidity scale. For a system check, identify the remote sensor as well. A result obtained by applying an electrical signal describes the display channel alone.
Suppose an indicator is intended to map 4–20 mA linearly to 0–100% RH. A 12 mA input should correspond to 50% RH on that configuration. This example checks the indicator scaling. It does not by itself test humidity sensing at the remote probe.
Specifications and service checks
A panel may be scaled 0–100 %RH while its input is a current or voltage from a remote transmitter. Copy both ends of the configured input span. Someone checking just the indicator can then reproduce the engineering-unit display without needing to guess the scaling.
- Range and configuration
- A display configured for 0–100 %RH may receive a voltage, current or digital signal. Its displayed scale should agree with the connected transmitter’s range.
- Manufacturers and models
- Vaisala HM40-series instruments illustrate an integrated RH indication; a panel indicator may instead use a separate transmitter.
- What we check
- For an indicator-only job compare the display with a known input. Include the humidity probe and an actual humidity comparison if a complete system result is required.
- Reference equipment and accuracy
- Use a traceable electrical source for an analogue input. Use a reference hygrometer when the sensor forms part of the calibration.
- When to repeat the check
- Recheck the display scaling whenever the transmitter is changed. Keep the electronic indicator’s history separate from humidity sensor drift.
Technical references Vaisala HM40 instrument guide · Vaisala HMP110 specifications