On this page
The interferometer kit or the machine axis?
Are you sending the interferometer for verification, or do you need a machine axis measured on site? List the module serial numbers. For angular and straightness measurements, include the optics fitted as well as the position-measuring kit. An earlier axis report often identifies that arrangement.
For an axis survey, the value written into the machine report is a length along a particular axis. For the interferometer kit itself, the result might instead concern a laser head, an air sensor or a set of angular optics. Name the required deliverable so the quote covers the kit calibration or the machine survey you need.
Laser interferometer configurations
List the attached temperature and pressure sensors separately. Indicate whether the required work concerns the reference system or measurements of a machine axis. A certificate for one sensor does not describe the entire optical system; agree which components, quantities and configurations the reported result will cover.
Specifications and service checks
A machine axis may be commanded to the same position from opposite directions. Keep the outward and return readings separate when comparing laser displacement with the controller. Averaging them immediately can conceal a reversal effect.
- Range and configuration
- Renishaw specifies linear measurement up to 80 m for the XL-80. Angular and straightness optics have their own ranges and resolutions.
- Manufacturers and models
- Renishaw XL-80 is a laser calibration system. Quote the laser head, compensation unit and optics used.
- What we check
- Compare the wavelength or indicated displacement with a traceable reference and identify the environmental compensation channels included in the service.
- Reference equipment and accuracy
- A traceable optical frequency or displacement reference establishes length. Air temperature, pressure and humidity sensors contribute uncertainty on long paths.
- When to repeat the check
- Review calibration after optical service or a damaged environmental sensor. Confirm performance before a critical long-axis survey.
Technical references Renishaw XL-80 laser system