Microplate washer testing

A microplate washer distributes and removes liquid at a row of positions. For equipment verification, dispensed volume and residual volume are distinct quantities. A single overall average may conceal a difference confined to one position on the manifold.

Send an enquiry
Microplate washer — equipment illustration
Equipment illustration
On this page

Locating an uneven row of deliveries

If a plate comes back with a wet strip down one side, show the affected wells on a sketch with A1 clearly marked. Note which end of the manifold faced A1 and whether the head was changed since the last check. The enquiry can then ask for results by position. A single figure for the volume delivered to the whole plate would hide the pattern you saw.

Microplate washer delivery and residual liquid

State whether the required evidence concerns dispensed volume, residual liquid, consistency between wells or another defined function. A passing reader check provides no evidence for the washer. RTS tests the washer under the specified plate, liquid and operating conditions.

Specifications and service checks

A recurring difference in one column can be hidden when all outlet volumes are added together. Identify the affected outlet and its position on the manifold if individual delivery checks are requested. Retain the plate orientation so that the laboratory and user refer to the same column.

Range and configuration
Agilent lists 96- and 384-well formats for the BioTek 405 TS. Plate format identifies the manifold configuration and is distinct from any separately measured volume indication.
Manufacturers and models
Agilent BioTek’s ELx808 is an absorbance reader and the 405 TS is a microplate washer.
What we check
For the service record, identify the reader’s optical channel or the washer’s separately measured indication. Link each calibration certificate to its serial number and fitted module.
Reference equipment and accuracy
For an optical indication, use a compatible certified optical reference with stated wavelength and uncertainty. A washer service record should identify any independently calibrated measuring channel.
When to repeat the check
Review after reader filter changes or replacement of a measuring module. Retain the reference identifier and date with the equipment record.