Muffle furnace calibration

A muffle furnace encloses the working space behind an internal chamber lining. A controller mounted outside this space measures through its own sensor. A comparison at the usual crucible position addresses the temperature experienced inside the chamber.

Send an enquiry
Muffle furnace — equipment illustration
Equipment illustration
On this page

Shelf position and changes to the lining

Identify the shelf, shelf height and normal operating points. If a new liner or element has been installed, keep the first subsequent temperature record with that work order. Comparing it with the earlier survey is more informative than comparing controller settings alone.

A shelf-level measurement made directly beneath the controller sensor may match the display even when another part of the chamber differs. If the process uses a single crucible position, specify that position. If several positions matter, show them on a sketch of the usable shelf.

Specifications and service checks

A metal tray sitting near the front of the chamber can occupy a different temperature zone from a small item placed deep inside. Mark the front edge and rear limit of the usual load. This gives the survey positions a practical relationship to the material actually heated.

Range and configuration
Carbolite Gero identifies the CWF 13/36 as a 1,300 °C chamber furnace. Separate the model’s maximum rating from its normal working temperatures and the dimensions of its useful zone.
Manufacturers and models
Carbolite Gero CWF laboratory furnaces and Nabertherm W-series industrial furnaces illustrate different chamber sizes and arrangements.
What we check
Compare independent temperature measurements with the furnace indication at the specified positions. For a zone survey, record the load arrangement and sensor locations.
Reference equipment and accuracy
Use a calibrated high-temperature probe and suitable readout rated for the muffle and temperature. The distance from the hot walls can matter when comparing a small chamber under light load.
When to repeat the check
Review after an element replacement, controller change or deterioration of the lining. A reference check at one position can flag a change between surveys.