Bath fluid and probe immersion
List the working temperatures and the particular fluid used at those settings. If you need several probes compared together, include their diameters and insertion depths. The results can then distinguish the bath display error from variation across the loaded working space.
A bath used for calibrating long probes and short probes may have two distinct working zones. Mark the positions and immersion depths actually used. A temperature difference with depth and a change over time at one position should be reported separately if both matter to the laboratory.
Specifications and service checks
When several reference probes share a bath, a shallow probe can sit nearer the fluid surface than the others. Record the sensing depth and oil level along with the working setpoint. This helps explain differences between positions even when the bath controller is steady.
- Range and configuration
- Fluke’s 6331 has a published working range of +35 to +300 °C. The permitted fluid and its condition can limit the usable span of an individual installation.
- Manufacturers and models
- Fluke Calibration’s 6331 is a deep-well hot bath. Its fluid requirements are part of the model specification.
- What we check
- Measure the oil temperature at the probe working depth and compare with the indication. If used for calibration, establish whether gradients and stability meet the intended work.
- Reference equipment and accuracy
- Use a traceably calibrated thermometer with a sheath and range suited to the bath. Reference corrections and immersion effects belong in the result.
- When to repeat the check
- Review performance after an oil change, pump repair or control adjustment. Trend the deviation at an established working point.
Technical references Fluke Calibration bath catalogue