Online TOC analyzer calibration

An online total organic carbon analyser can report a local value and send another value to a plant historian. A service request should identify whether the analyser response, the transmitted output or both require verification. Their scaling and units can differ.

Send an enquiry
Online TOC analyzer — equipment illustration
Equipment illustration
On this page

The installed sampling point and transmitted reading

Record the sample-point tag and the location of any reference connection specified in your site procedure. Keep display readings and transmitted values with their timestamps. The time stamps and output scaling can explain a difference between the local reading and the value saved by the plant system.

Online TOC analyzer and sample path

Identify its inlet conditions, sampling arrangement and any signal passed to a control system. Specify whether the request concerns the analyzer reading alone or also the transmitted value. Agreement on that boundary prevents an instrument comparison from being filed as evidence for the entire water-system monitoring installation.

Specifications and service checks

A reading in the plant historian may be delayed by sample transport or the instrument’s reporting interval. If it differs from a local reading, save both timestamps and the selected units. This lets the service team separate a timing or scaling question from a concentration response check.

Range and configuration
Shimadzu’s TOC-4200 has full-scale ranges from 0–5 through 0–1,000 mg C/L, with a higher-range dilution option. State which range is active in the installed configuration.
Manufacturers and models
Shimadzu’s TOC-L series is a benchtop example and the TOC-4200 is an online analyser.
What we check
Compare the reported carbon concentration with suitable certified reference solutions across the range actually used. Identify the reported fraction, instrument mode and any installed dilution option.
Reference equipment and accuracy
Use carbon standards traceable in the stated concentration units with uncertainties suited to the acceptance limit. Blanks and low-level reference checks matter near the lower end.
When to repeat the check
Review after changes to the detector, sampling path or dilution equipment. Use reference-material history to decide when the interval needs to change.

Technical references Shimadzu TOC-L specifications · Shimadzu TOC-4200 specifications