Optical pyrometer calibration

An optical pyrometer may use a visual brightness match or an electronic optical detector. The model and viewing arrangement are needed to specify the check. A device with a sighting telescope is not necessarily a visual matching instrument.

Send an enquiry
Optical pyrometer — equipment illustration
Equipment illustration
On this page

Working distance and the optical path

State the temperature range, sighting distance and any optical window used on the job. If the operator applies an emissivity setting or looks through an attenuation filter, keep that detail with the result for comparison with earlier records.

If an operator sights through a window, include the window and its condition in the description of the installed measurement. Identify whether the request concerns the pyrometer alone or the complete sight path. Results obtained against a direct reference target refer to the stated optical arrangement.

Specifications and service checks

A disappearing-filament instrument and an electronic optical instrument may both be called optical pyrometers on an asset list. Identify the viewing system and range actually used. If the operator observes a target through a window, record the window and viewing conditions with that range.

Range and configuration
The Spectrodyne DFP 3000 has four overlapping ranges. Its first covers 775–1,225 °C and its second covers 1,075–1,760 °C. Specify the range used for readings within the overlap so the report identifies the setting.
Manufacturers and models
Spectrodyne’s DFP 3000 is a visual, disappearing-filament instrument. The operator matches the brightness of a calibrated lamp filament to the target. Electronic optical pyrometers read a detector instead.
What we check
Compare brightness matches at assigned radiance temperatures with the fitted filter and range identified. Repeated matches give a measure of the observer’s repeatability as well as the scale correction.
Reference equipment and accuracy
The blackbody aperture must fill the pyrometer’s field of view. Select a source whose radiance-temperature uncertainty is suitable at the intended high-temperature points.
When to repeat the check
Lens contamination and changes to a sight tube can affect an installed result even if the electronics remain stable. Check after optical work and use trends to set the recall.

Technical references AMETEK Land SPOT+ specifications · Fluke infrared thermometer calibration guide · Spectrodyne DFP 3000 visual pyrometer