+91 70565 55582[email protected]Get a quoteRequest a call
RTS Calibration
Menu

Orbital shaker calibration

An orbital shaker moves its platform in a circular path while the platform keeps its orientation. Shaking speed and orbit diameter are different characteristics. The same speed indication can refer to substantially different movement on two models with different orbits.

Send an enquiry
Orbital shaker — equipment illustration
Equipment illustration
On this page

Speed, orbit and elapsed time

Give us the platform, load and settings used on the bench. We can check speed in cycles per minute, and the diameter of the orbit if your procedure calls for it. For a timed programme, include the elapsed time as a third reading.

Two platforms may run at the same orbits per minute but have different orbit diameters. If your comparison is specifically about the motion, include the orbit diameter in the order. A timing comparison on a five-minute programme and a measurement of the steady running rate are separate observations.

Orbital shaker motion and loaded platform

State the speed settings, load condition and timing requirements used by the method. Orbital diameter and rotations per minute describe different aspects of motion. If a temperature-controlled enclosure is fitted, list its temperature requirements separately instead of treating one speed result as a complete qualification.

Specifications and service checks

Two shakers can both indicate 150 rpm but have different orbit diameters. If a process was transferred between them, list the orbit and platform attachment as well as the speed. The speed comparison alone cannot describe the difference in platform motion.

Range and configuration
IKA specifies a 10 mm orbit for the KS 260 basic and a maximum speed of 500 rpm. Orbit size and orbital speed are separate characteristics.
Manufacturers and models
IKA supplies KS orbital shakers and EUROSTAR overhead stirrers. The motion type on the nameplate is a useful first distinction.
What we check
Compare the indicated motion with a reference speed measurement at the working points. Specify shaft rotation or orbital cycles and identify the fitted platform or attachment.
Reference equipment and accuracy
Use a calibrated tachometer, strobe or cycle counter that suits the observed movement. A calibrated timer supports any separate run-time comparison.
When to repeat the check
A worn coupling or drive service can change the relationship between indication and movement. Review speed at a familiar setting between scheduled visits.

Technical references IKA EUROSTAR 20 digital technical data · IKA KS 260 basic technical data

Enquire about Orbital shaker

Service enquiry