The input span or electrode connection
For a receiving indicator, supply the configured input span and low and high displayed values. For a direct electrode input, identify the reference electrode arrangement. Keeping these configurations separate helps explain a reading that differs between the plant display and a handheld meter.
For a panel instrument fed from a separate transmitter, identify whether its input is a millivolt signal or a scaled analogue signal. A reading checked at the indicator terminals establishes the operation of that channel. Record the transmitter or electrode separately when the full loop is included.
Specifications and service checks
A panel set up to show millivolts can also have an offset entered in the configuration menu. Copy that setting before the service visit. Comparing the displayed value with an applied millivolt input will then account for the adjustment the process operator actually sees.
- Range and configuration
- Hanna lists ±1,000 mV for its HI98121. The electrode system and the temperature associated with an assigned redox solution value matter when interpreting an observed potential.
- Manufacturers and models
- Hanna’s HI98121 is a combined pH and ORP instrument with a millivolt indication.
- What we check
- If the panel receives a scaled transmitter signal, compare that electrical input with the displayed value. A direct redox electrode input requires a different electrical reference and configuration.
- Reference equipment and accuracy
- Use a traceable low-voltage reference for the input, and an identified chemical reference with a value appropriate to the measurement temperature for the electrode system.
- When to repeat the check
- Follow solution checks between electronic calibrations. Replace or investigate an electrode showing unstable or displaced readings, then document the new assembly.
Technical references Hanna HI98121 specifications