ORP meter calibration

An oxidation-reduction potential meter reports a voltage relative to a reference electrode. A reference solution and a simulated electrical input check different parts of the system. Identify the electrode used in operation and the form of check required.

Send an enquiry
ORP meter — equipment illustration
Equipment illustration
On this page

The reference solution and its temperature

Record the reference solution identification, temperature and expected value under its stated conditions when assessing the complete meter. For an electronic-only comparison, retain the applied millivolts and display reading. A result from one check should be recognisable as such in the instrument history.

When comparing two ORP readings, note the reference electrode in each assembly and the temperature at the time of measurement. The displayed millivolt value belongs to that combination. A disagreement between two field meters cannot be traced confidently from the numerical readings alone.

Specifications and service checks

A redox reading is conventionally reported against a reference electrode. If a replacement probe gives a different value in the same reference solution, identify the two probe types and the solution temperature. This gives context for any difference in the electrode arrangements.

Range and configuration
Hanna lists ±1,000 mV for its HI98121. The electrode system and the temperature associated with an assigned redox solution value matter when interpreting an observed potential.
Manufacturers and models
Hanna’s HI98121 is a combined pH and ORP instrument with a millivolt indication.
What we check
Check the electronics with a suitable millivolt source. For the complete measurement chain compare the meter and electrode with an assigned redox reference solution.
Reference equipment and accuracy
Use a traceable low-voltage reference for the input, and an identified chemical reference with a value appropriate to the measurement temperature for the electrode system.
When to repeat the check
Follow solution checks between electronic calibrations. Replace or investigate an electrode showing unstable or displaced readings, then document the new assembly.

Technical references Hanna HI98121 specifications