Source, simulation and measurement routines
Describe one representative task per function. For a simulated temperature sensor, identify the sensor type and wiring. For a current loop, specify whether the unit measures, sources or simulates. Include the signal reported by the calibrator separately from the response of the process instrument.
When a calibrator displays 100 °C in RTD simulation mode, its output is an electrical resistance representing a selected sensor curve. That reading is not a temperature generated in a block or bath. Identify the curve and input seen by the receiving controller.
Specifications and service checks
A process calibrator may source a thermocouple signal at the bench but read a transmitter loop in the field. If both duties appear under one asset number, list the temperature simulation and the current input separately. They exercise different functions.
- Range and configuration
- Record the electrical input or output, sensor table and polarity. A displayed temperature can represent a simulated resistance or a thermocouple voltage.
- Manufacturers and models
- Fluke 725 is a portable multifunction process calibrator. Its temperature simulation, loop and voltage functions require different reference arrangements.
- What we check
- Compare measurement and source modes independently for the functions used. Include the reference-junction mode for thermocouple work and lead configuration for RTDs.
- Reference equipment and accuracy
- Use a traceable electrical calibrator, reference meter, resistance standards and appropriate temperature reference as the configuration requires.
- When to repeat the check
- Trend the functions carried into the field most often and review after a terminal or module change.
Technical references Fluke 725 Multifunction Process Calibrator