On this page
Detector type and reference field
Provide the detector type and the radiation quality relevant to the monitoring programme. Include the range, units and previous certificate for that pairing. An observed count rate and a displayed dose rate are different quantities and should retain their own labels in the report.
A replacement detector head may arrive with its own calibration certificate. Your instrument register also needs the reader or ratemeter used with it, particularly when the reader applies a stored factor. Note the factor reference and date so the review covers the pairing actually in use.
Radiation detector and readout pairing
List the radiation quality and response quantity needed, using the terminology in the previous certificate. Replacing a detector or connecting it to a different meter changes the assembly under review, even when the connector fits without an adapter.
Specifications and service checks
A removable detector head can have its own response range. Include both the head and the readout serial numbers when booking the job, along with the quantity and unit the display was set to show.
- Range and configuration
- A count rate in counts per minute and a dose rate in sieverts per hour are different quantities. Conversion depends on detector response and radiation quality.
- Manufacturers and models
- Ludlum’s Model 44-9 is a detachable alpha, beta and gamma detector.
- What we check
- Check the probe with the meter and cable identified for use. An authorised laboratory determines response for the requested radiation and measurement quantity.
- Reference equipment and accuracy
- Use traceable radiation references and documented geometry under a qualified radiation programme. Match the calibration to the probe and meter combination.
- When to repeat the check
- Recheck after a detector window, cable or electronics change. Retain the probe serial number alongside the meter record.
Technical references Ludlum Measurements product reference