Radiation monitor testing

A fixed area monitor may show its reading locally and send an alarm to another panel. Checking whether that panel receives an alarm establishes the response of the alarm path. To check the accuracy of the displayed quantity, ask for a comparison on the monitoring channel itself.

Send an enquiry
Radiation monitor — equipment illustration
Equipment illustration
On this page

Local indication, remote display and alarms

List the installed detector, display units, remote channel and alarm functions covered by your radiation protection programme. For a networked installation, give the channel designation used by the control room. The service record can then distinguish a detector response from a downstream indication issue.

Radiation monitor at a fixed location

Identify each channel and the location where its response is required. Distinguish detector calibration from alarm-chain verification and a survey of the room. Supply the owner’s controlled procedure so the proposed specialist work has a defined boundary and a responsible technical reviewer.

Specifications and service checks

Compare the label on a saved trend with the live display unit. A monitor can export an accumulated value while the panel shows a rate. The time interval and quantity identify which figures can be compared.

Range and configuration
The controller’s indication and alarm span depend on its configuration and detector. Include the displayed units and the alarm channels in the request.
Manufacturers and models
The Ludlum Model 375 is an area monitor controller used with a compatible detector.
What we check
Distinguish a check of the controller indication and contacts from a response check of the installed detector in its radiation field.
Reference equipment and accuracy
Use calibrated signal references for the controller and authorised field references for the detector. Document which parts of the installed system the results cover.
When to repeat the check
Review calibration after replacing a detector or altering its configuration. The radiation protection programme sets routine tests of indication and alarms.

Technical references Ludlum Measurements product reference

Enquire about Radiation monitor