RTD calibrator calibration

An RTD calibrator may simulate a resistance thermometer to drive a controller or measure an actual RTD. Both functions may display temperature even though the electrical comparison is made through resistance.

Send an enquiry
RTD calibrator — equipment illustration
Equipment illustration
On this page

Simulating an RTD or reading a probe?

A three-wire Pt100 input on a controller and a two-wire input on a panel meter need different connections even when both are set to the same temperature. Mark the RTD type, resistance curve and number of wires at the points your shop uses. If the calibrator also reads a probe directly, include the measurement mode as another function on the order.

A Pt100 designation establishes nominal resistance at 0 °C, but a simulator may offer several curves. Agreement near zero alone does not establish agreement elsewhere. Record the curve and the resistance or temperature points actually used in the process.

Specifications and service checks

For a simulated Pt100, the plant input may use two, three or four leads. Identify which wiring arrangement it expects. A comparison at the simulator’s terminals and a temperature indication through the installed lead arrangement do not include exactly the same influences.

Range and configuration
A Pt100 has nominal resistance of 100 Ω at 0 °C. A Pt1000 has 1,000 Ω at that point, so confusing the tables changes the electrical input by a factor of ten.
Manufacturers and models
Fluke 712B measures and simulates resistance temperature detectors. Record the sensor curve and lead arrangement selected for each function.
What we check
Compare simulated resistance with a calibrated readout and check measuring mode with assigned resistance values for the selected temperature points.
Reference equipment and accuracy
Use calibrated resistance standards or a resistance simulator and a suitable reference ohmmeter. Lead resistance matters in two-wire use.
When to repeat the check
Repeat after terminal repair or a sensor-table change, and use a stable resistance checkpoint between calibrations.

Technical references Fluke 712B RTD Calibrator

Enquire about RTD calibrator