Matching energy registers to current inputs
Specify the register identifiers, phase arrangement, current inputs and tariff or import/export direction used for billing checks. If an optical pulse, display and downloaded register are compared, identify each result separately and note the meter configuration in force during the test.
An export register on the display and an import register in the data logger can both be reading correctly while showing different totals. Note the register identifier beside each reading before investigating a possible energy measurement error.
Specifications and service checks
For a time-of-use dispute, distinguish total imported energy from the allocation to tariff registers. Correct accumulated kWh can still be assigned to the wrong period if the clock or tariff schedule is different. Supply the register names and the relevant time window.
- Range and configuration
- The register records kWh; a pulse output may be specified in impulses per kWh. Test current and power factor influence error even when the nominal voltage is fixed.
- Manufacturers and models
- Schneider Electric's iEM3200 series includes transformer-connected energy meters such as the iEM3255. Identify the programmed CT ratio and pulse constant.
- What we check
- Compare reference energy against the billing register or pulse output, recording meter constant, tariff register and CT ratio. If a network reading is used, confirm its unit and scaling against the local indication.
- Reference equipment and accuracy
- Use a calibrated energy reference or power standard with traceable integration time. Current transformer error belongs in the result only when that transformer is included.
- When to repeat the check
- For revenue metering, confirm the required verification regime. Review the settings and result after a CT ratio or firmware change.
Technical references Energy meters iEM3100 / iEM3200 / iEM3300 series User manual