On this page
The detector and meter as a pair
Identify the detector and meter as a pair and state the measurement ranges in the site's monitoring programme. Record alarm operation separately if it is part of the request. If several probes fit the same meter, give each probe and meter pairing its own entry on the service record.
A personal monitor and a hand-held survey meter may both display a dose-rate unit. If the survey instrument is used with an extension probe, identify whether the extension cable and probe were fitted for the calibration. That detail can be easy to lose when only the serial number on the meter is copied to the register.
Survey meter range and radiation response
Include the radiation type and energy conditions relevant to its intended work, rather than requesting only a general meter certificate. A result obtained with one probe or radiation quality cannot automatically establish response under every field condition. The report identifies the radiation quality, probe and selected range used for each comparison.
Specifications and service checks
A survey log may export 0.005 mSv/h while the handheld display shows 5 µSv/h. These are equivalent dose rates. Check the unit prefix and the stored time before assigning an apparent thousandfold change to the detector.
- Range and configuration
- A meter and probe can indicate count rate or a calibrated radiation quantity, depending on the scale, probe and calibration. Supply the scale legend and switch positions.
- Manufacturers and models
- Ludlum’s Model 3 is a general-purpose survey meter with selectable ranges. Mirion’s ADM300A is another survey meter family.
- What we check
- Have the meter and its assigned probe assessed as a pair for the measurement quantity used at the site. Record the response on each range requested.
- Reference equipment and accuracy
- A qualified radiation laboratory selects traceable sources or fields and the appropriate reference detector. A pulse check of the meter tests the electronics alone.
- When to repeat the check
- Recheck the pairing when a probe changes. Keep routine response checks and calibration certificates against both serial numbers.
Technical references Ludlum Measurements product reference · Mirion product reference