Tablet hardness tester qualification

A tablet hardness tester reports the force at which a test piece breaks. Some instruments also report thickness, diameter or mass. A calibrated force indication and a dimensional reading are separate quantities, even if they share the same display.

Send an enquiry
Tablet hardness tester — equipment illustration
Equipment illustration
On this page

Jaw arrangement and breaking force

A tester used on several tablet shapes may have more than one jaw set. Give the jaw arrangement and the force range you use for each shape. If an old worksheet uses kgf and the new display reads N, flag that before comparing the numbers. Thickness or diameter, where fitted, should appear under its own unit alongside the force results.

Tablet hardness tester breaking-force function

The record identifies the force unit, operating mode and any dimensional channels used by the laboratory. Model-specific references and contact arrangements are retained with the results. Tablet breaking-force measurement and metal hardness testing use different measured quantities and reference methods, which are identified in their respective procedures.

Specifications and service checks

If a tablet tester is compared with an independent force device, record whether its display is set to newtons or another force unit. A numeric change after resetting the display can be a unit change. Also identify the loading direction used by the fitted jaws.

Range and configuration
ERWEKA gives a standard 10–300 N force range for its TBH 125 and a 10–500 N extended option. Confirm which version is fitted before requesting upper-range points.
Manufacturers and models
ERWEKA’s TBH 125 data sheet lists its standard and extended force ranges separately.
What we check
Compare indicated force against an independent force reference in the instrument’s loading direction and over the relevant part of the range.
Reference equipment and accuracy
Use a calibrated force transducer or suitable manufacturer-approved force reference with uncertainty adequate for the permitted force error. The reference fixture and alignment matter.
When to repeat the check
Review after a load-cell change, overload or jaw repair. Follow the force comparison history when setting the next service interval.

Technical references ERWEKA TBH 125 data sheet