Thermocouple calibrator calibration

A thermocouple calibrator may generate a signal to test an indicator, measure the signal from a probe, or do both. State the thermocouple types and the operating mode for each requested point. The source and measure functions are separate checks.

Send an enquiry
Thermocouple calibrator — equipment illustration
Equipment illustration
On this page

Reference junctions and simulated temperature

Copy the reference-junction setting used at the bench and identify any external junction or adapter. Where results are used in millivolts as well as temperature, ask for the output units to appear with the instrument's type setting.

If a calibrator is checked as a millivolt source but normally used on site to simulate a type K thermocouple, the required temperature conversion and reference-junction arrangement need separate consideration. Mark the mode used by the technician rather than specifying only the electrical range printed on the case.

Specifications and service checks

A calibrator sourcing thermocouple voltage can use internal reference-junction compensation or a specified external reference. Copy that setting from both instruments before comparing the displayed temperatures. A change of compensation convention can create an apparent difference even when the voltage is reproduced accurately.

Range and configuration
Fluke lists 17 thermocouple types plus millivolts on the 714B. Each type has its own temperature-to-voltage relationship and usable span.
Manufacturers and models
Fluke’s 714B is a dedicated thermocouple calibrator.
What we check
Compare the electrical source and measure modes separately. Record the reference-junction setting and connector arrangement for results stated in temperature units.
Reference equipment and accuracy
A traceable low-voltage standard checks the electrical signal. A calibrated thermometer or suitable junction reference is needed where the comparison includes junction compensation.
When to repeat the check
A replacement input socket or a significant change in junction error warrants a check. Trend the readings for the types actually used.

Technical references Fluke 712B and 714B specifications · EURAMET thermocouple calibration guide

Enquire about Thermocouple calibrator