Peak capture and the fitted torque sensor
List the input channel or fitted sensor, units, direction and acquisition mode. Include the joint simulator if it is part of the line check. For stored readings, note the filter or reset setting used before a value is accepted.
If the quality log contains the first peak after a driver starts turning, a maximum taken later in the tightening cycle may be a different value. Identify the peak definition used on the production log when preparing the order.
Specifications and service checks
A powered driver can give several peaks in one fastening cycle. Keep a short trace or identify the analyser’s peak-selection setting when comparing a workshop printout with the driver’s stored torque. The highest peak may occur at a different event.
- Range and configuration
- On a modular analyser the torque range follows the attached transducer. The Norbar T-Box 2 manual illustrates the point with a 500 N·m sensor on one channel.
- Manufacturers and models
- Norbar’s T-Box 2 is a modular analyser with two transducer inputs.
- What we check
- Compare peak or steady readings against applied reference torque in the operating direction. If a tool release event is part of the test, identify the capture mode and its settings.
- Reference equipment and accuracy
- Use a traceable torque transfer standard or a known force and effective lever length. Check that the reference, sensor and indicator all cover the requested lowest point.
- When to repeat the check
- Check again after changing the transducer, sensor scaling or peak capture firmware. Review the previous comparison when deciding the next interval.
Technical references Norbar T-Box 2 manuals and software