Temperature along the length of the tube
Mark axial distances from an identifiable end or flange. When several operating temperatures are required, retain the position coordinates for every run so the report can show where the usable uniform length begins and ends.
If the working fixture occupies the length from 50 to 150 mm from an end flange, identify that interval on the survey request. A result at the geometric centre of the tube alone does not show how much of the fixture lies inside the required temperature band.
Specifications and service checks
A long work tube can project beyond both ends of its heated section. Mark the axial positions occupied by the parts and distinguish them from the overall tube length. This defines the length over which a temperature profile would actually be useful.
- Range and configuration
- The Carbolite Gero TF1 12/125/600 has a maximum rating of 1,200 °C, a 125 mm furnace bore and a 600 mm heated length. The uniform working zone is a separate dimension that depends on the setup.
- Manufacturers and models
- Carbolite Gero lists single- and multiple-zone arrangements in its tube-furnace product literature. The TF range specifies heated length and bore.
- What we check
- Compare temperatures along the agreed length with a traceable probe. Identify the work tube, axial positions and any end fittings present during the survey.
- Reference equipment and accuracy
- Use a calibrated thermocouple or suitable reference thermometer rated for the temperature. Position error matters where the axial gradient is steep.
- When to repeat the check
- Revisit the map after replacing a work tube, elements or control sensor. Track both centre-point drift and changes near the zone boundaries.
Technical references Carbolite Gero TF tube furnaces