On this page
Tank positions and the specified performance check
Identify the tank, basket and work positions referenced by your approved performance method. Where the requirement is for a qualitative test, preserve the observations and acceptance criterion. Where a numeric result is required, specify the quantity and reference beforehand.
After a basket is replaced, the tank tag may be unchanged but a previous result can describe a different arrangement. Note the basket reference and mark the occupied working area. If the report also needs the heater and timer checked, give their usual indications and which operation the timer measures.
Ultrasonic cleaner timer and bath conditions
For an ultrasonic cleaner, a bath temperature or timer reading tells you little about ultrasonic activity or cleaning performance. If those are the quantities you need assessed, give the test method and load specified in your procedure. We can then record the performance result separately from the temperature and time readings.
Specifications and service checks
A bath timer ending on cue and a correct display temperature establish two specific functions. If the enquiry concerns cleaning action as well, identify the separate functional assessment required by your department. A timer comparison does not measure the acoustic output in the tank.
- Range and configuration
- Elma lists a nominal ultrasonic frequency of 37 kHz for the Elmasonic Select series. This identifies the drive frequency. Its bath-temperature and timer indications have separate ranges and checks.
- Manufacturers and models
- Elma’s Elmasonic Select family includes tanks in several sizes.
- What we check
- For temperature or timing, compare the display with an independent reference. Document any separate functional assessment against a defined user or manufacturer requirement.
- Reference equipment and accuracy
- Use a calibrated thermometer or time reference for the corresponding channel. An electrical power reading alone cannot establish the acoustic field within the bath.
- When to repeat the check
- Recheck after heater, timer or transducer service as appropriate. Keep the instrument checks separate from routine condition and cleaning records.
Technical references Elma Elmasonic Select catalogue · Elma Elmasonic Select technical specifications