What does the display show when vented?
When the chamber is vented, record the displayed value and its unit. Include the sensing head, controller channel and the pressure range used during pump-down. If the indicator has been replaced, give us both component identities so the older certificate can be matched to the right assembly.
A display marked 95 percent vacuum and another marked in mbar absolute can refer to the same process under particular atmospheric conditions. Preserve the absolute reference or assumed ambient pressure if one reading will be converted to compare with the other.
Specifications and service checks
A local vacuum display may keep showing the last sound reading when a sensor cable is disconnected, or it may issue an overrange or fault indication. If the concern is a frozen panel value, note the displayed status symbol together with the number and the sensor connection.
- Range and configuration
- A controller can display many pressure decades because it accepts several head types. The installed head determines the usable range and gas response.
- Manufacturers and models
- Leybold GRAPHIX controllers read compatible vacuum gauge heads. Identify both reader and head when ordering a system check.
- What we check
- Compare the reading of the connected head against known vacuum pressures. An input simulation is useful when only the electronics are under examination.
- Reference equipment and accuracy
- Use traceable vacuum references across the selected decades, or a calibrated signal source for an electronics-only check.
- When to repeat the check
- Review after head exchange or contamination. Store the serial number of each head beside its controller channel.