Shelf temperature at the working pressure
For a temperature survey, identify the shelf locations and the vacuum condition at which the results are needed. A comparison made with the door closed at atmospheric pressure does not describe the same heat-transfer conditions as operation under vacuum.
A vacuum reading recorded simply as 100 mbar is ambiguous if one document uses absolute pressure and the other reports a change relative to room pressure. Preserve the pressure reference and shelf location with each temperature survey so earlier and later runs can be compared meaningfully.
Specifications and service checks
A shelf probe in contact with a tray and a probe suspended in the chamber see different thermal conditions under vacuum. Mark which position represents the load and record the absolute pressure used for the observation. The controller setpoint alone cannot identify the arrangement.
- Range and configuration
- BINDER states +9 °C above ambient to +220 °C for its VD 56. The shelf-temperature range and the chamber’s absolute-pressure range need separate entries on the request.
- Manufacturers and models
- BINDER’s VD 56 is a vacuum drying chamber with conducting shelves.
- What we check
- Specify which shelf and pressure condition apply to the temperature comparison. If the vacuum indication matters, request a separate pressure comparison at the relevant absolute pressures.
- Reference equipment and accuracy
- Use calibrated contact temperature probes suitable for the shelf and an absolute-pressure reference for the gauge. Contact arrangement matters under vacuum.
- When to repeat the check
- Recheck affected quantities after a shelf sensor, pressure gauge or door-seal change. The two channels can have different recall intervals.
Technical references BINDER VD 56 specifications