On this page
Distinguishing pump performance from the system reading
Identify the pump and the system on which the observation was made. For an enquiry about pump-down time or ultimate pressure, give the chamber volume, pressure interval and gauge used. List the pressure gauge separately if a calibrated indication is also required.
Vacuum pumps in a calibration setup
Record the pump type, connections and intended use. If the requirement concerns ultimate pressure, pumping behaviour or another performance quantity, provide the method and conditions for review. RTS records pump performance under the specified operating conditions separately from calibration of the pressure gauge.
Specifications and service checks
For a pump taking longer to evacuate a fixture, give the chamber volume, starting pressure and target pressure alongside elapsed time. A blanked-off inlet check and a pump-down time measured on the loaded fixture answer different maintenance questions.
- Range and configuration
- State the quoted pumping speed and the ultimate pressure required in the intended operating condition. A system's leaks and gas load can dominate the installed result.
- Manufacturers and models
- Edwards RV8 is a two-stage rotary vane vacuum pump. Pump speed and ultimate pressure refer to different performance quantities.
- What we check
- Verify ultimate pressure under agreed conditions. Where pumping speed is required, define the gas throughput and pressure at which speed is reported.
- Reference equipment and accuracy
- A calibrated vacuum reference is needed at the test port. A throughput or flow reference is also required for pumping-speed measurement.
- When to repeat the check
- Consider a performance check after service or a change in evacuation time. Maintain the vacuum reference on its own calibration schedule.
Technical references Edwards RV rotary vane pump specifications