Switching direction and pressure reference
Give the switching direction in words, the contact in use, units and pressure reference. Include the trip and recovery settings separately. If a controller delays the alarm, specify whether timing of that separate controller is part of the requested check.
For example, a switch that trips at 50 mbar absolute during evacuation might reset at 70 mbar absolute as air is admitted. Record both pressures with the contact state. This makes the switching difference visible even if the maintenance log calls the first movement “vacuum rising”.
Specifications and service checks
A switch described as “on at vacuum” might energise a lamp through a normally closed contact that opens as pressure falls. Mark the terminal pair and the required machine action. This removes ambiguity when the service team records the transition and reset.
- Range and configuration
- Specify the trip pressure and reset pressure as absolute pressure or vacuum relative to atmosphere. The two descriptions are not interchangeable.
- Manufacturers and models
- Leybold PS 115 is an example of a vacuum pressure switch. Record the required operating direction and contact state.
- What we check
- Compare the pressure at switching during evacuation and during recovery. Repeat at the point used to control the process.
- Reference equipment and accuracy
- Use a calibrated gauge for the relevant vacuum range and a contact monitor. Low-end resolution should be sufficient to distinguish the switch differential.
- When to repeat the check
- Review after adjusting the threshold, changing the switch or observing erratic switching.
Technical references Leybold vacuum switch catalogue