Water bath calibration

A water bath used for comparisons needs a defined working area below the liquid surface. Mark the usual fill level, rack position and probe depth. Temperatures at the wall, in the centre and close to the surface can differ.

Send an enquiry
Water bath — equipment illustration
Equipment illustration
On this page

Comparing the water and the load

For a display comparison, request a reading at a specified reference position. For a bath used to hold many instruments at once, ask for a survey across their actual positions and identify whether the circulation system was operating during the measurements.

For a bath used with a rack of sample tubes, describe whether the check concerns the water or the temperature inside an identified vessel. A reading from water beside the tube and one inside it can settle at different rates, especially after the rack is loaded.

Specifications and service checks

If a bottle is held in a rack, the water temperature beside it and the temperature inside it can settle at different times. State which quantity your method requires and mark the water level and rack position. This defines the intended comparison before the bath is mapped.

Range and configuration
The Memmert WTB15 controller can be set from +10 to +100 °C. The lower achievable operating temperature also depends on ambient temperature and any fitted cooling accessory.
Manufacturers and models
Memmert’s WTB15 is a laboratory water bath; JULABO makes circulating baths with a different flow arrangement.
What we check
Compare the measured water temperature at selected rack positions with the controller display. Distinguish position-to-position variation from drift over time.
Reference equipment and accuracy
Use calibrated immersion thermometers or suitable waterproof probes. Their uncertainty and time response must suit the tolerance applied to the bath.
When to repeat the check
Check after a change to the rack, heater or circulator. For routine recall use previous gradients and the results at the normal working setting.

Technical references Memmert WTB15 specifications · JULABO CORIO CD-200F specifications